Hanna Instruments HI 9026 Instruction Manual Download Page 8

8

For accurate ORP measurements, the surface of the electrode must be

clean and smooth. Pretreatment solutions are available to condition

the electrode and improve its response time (see "Accessories" section).

Notes:

• When the reading is out of range, the display will flash the closest

full-scale value.

• If using pH electrode while in mV mode, the meter will measure

the mV generated by the pH electrode.

MEM & MR FUNCTIONS

From normal measurement mode, pressing the MEM key will freeze

the readings on the display and store into internal memory the

current values (pH and temperature, or mV and temperature)

together with “condition” and buffer segments.

The MEM tag blinks and the display remains frozen until the MEM

key is pressed again.

Note

: While in MEM mode, the user can switch between pH and mV

by pressing the RANGE key.

A stored value can be recalled by pressing MR: the display will show

the value, together with the RCL & MEM tags, as long as the MR key

is pressed.

Note

: Pressing MR, only the range that was active at the time MEM

was pressed is shown.

BACKLIGHT FEATURE

The meter is provided with a Backlight feature,

which can be easily toggled on and off through

the keyboard by pressing the LIGHT key.

Note

: The backlight automatically shuts off after approximately 1 minute

with no buttons pressed.

21

ACCESSORIES

pH CALIBRATION SOLUTIONS

HI 70004P

pH 4.01 Buffer Solution, 20 mL sachet, 25 pcs

HI 70006P

pH 6.86 Buffer Solution, 20 mL sachet, 25 pcs

HI 70007P

pH 7.01 Buffer Solution, 20 mL sachet, 25 pcs

HI 70009P

pH 9.18 Buffer Solution, 20 mL sachet, 25 pcs

HI 70010P

pH 10.01 Buffer Solution, 20 mL sachet, 25 pcs

HI 7001L

pH 1.68 Buffer Solution, 500 mL bottle

HI 7001M

pH 1.68 Buffer Solution, 230 mL bottle

HI 7004L

pH 4.01 Buffer Solution, 500 mL bottle

HI 7004M

pH 4.01 Buffer Solution, 230 mL bottle

HI 7006L

pH 6.86 Buffer Solution, 500 mL bottle

HI 7006M

pH 6.86 Buffer Solution, 230 mL bottle

HI 7007M

pH 7.01 Buffer Solution, 500 mL bottle

HI 7007M

pH 7.01 Buffer Solution, 230 mL bottle

HI 7009L

pH 9.18 Buffer Solution, 500 mL bottle

HI 7009M

pH 9.18 Buffer Solution, 230 mL bottle

HI 7010L

pH 10.01 Buffer Solution, 500 mL bottle

HI 7010M

pH 10.01 Buffer Solution, 230 mL bottle

ELECTRODE STORAGE SOLUTION

HI 70300L

Storage Solution, 460 mL bottle

HI 70300M

Storage Solution, 230 mL bottle

ELECTRODE CLEANING SOLUTIONS

HI 70000P

Electrode Rinse Solution, 20 mL sachet, 25 pcs

HI 7061L

General Cleaning Solution, 460 mL bottle

HI 7061M

General Cleaning Solution, 230 mL bottle

HI 7073L

Protein Cleaning Solution, 460 mL bottle

HI 7073M

Protein Cleaning Solution, 230 mL bottle

HI 7074L

Inorganic Cleaning Solution, 460 mL bottle

HI 7074M

Inorganic Cleaning Solution, 230 mL bottle

HI 7077L

Oil & Fat Cleaning Solution, 460 mL bottle

HI 7077M

Oil & Fat Cleaning Solution, 230 mL bottle

REFILLING ELECTROLYTE SOLUTIONS (50 mL, 4 pcs)

HI 7071

3.5M KCl+AgCl Electrolyte for single junction electrodes

HI 7072

1M KNO

3

 Electrolyte

HI 7082

3.5M KCl Electrolyte for double junction electrodes

HI 8093

1M KCl+AgCl Electrolyte

ORP PRETREATMENT SOLUTIONS

HI 7091L

Reducing Pretreatment Solution, 460 mL bottle

HI 7091M

Reducing Pretreatment Solution, 230 mL bottle

HI 7092L

Oxidizing Pretreatment Solution, 460 mL bottle

HI 7092M

Oxidizing Pretreatment Solution, 230 mL bottle

Summary of Contents for HI 9026

Page 1: ...el 21 4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 662 619 0708 11 Fax 662 619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765...

Page 2: ...ts prepaid When shipping any instrument make sure it is properly packed for complete protection TABLE OF CONTENTS WARRANTY 2 PRELIMINARY EXAMINATION 3 GENERALDESCRIPTION 3 FUNCTIONAL DESCRIPTION 4 SPECIFICATIONS 5 OPERATIONAL GUIDE 6 pH CALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 13 SETUP MENU 14 mV CALIBRATION 15 TEMPERATURE CALIBRATION 15 BATTERYREPLACEMENT 15 LCDMESSAGEGUIDE 16 TEMPERATURE CORR...

Page 3: ...irely new dimension to the measuremeny of pH by allowing the user to dramatically improve the reliability of the measurement 7 memorized buffers pH 1 68 4 01 6 86 7 01 9 18 10 01 and 12 45 for calibration Custom calibration by entering up to two custom buffers Messages on the LCD to make the calibration easy and accurate Diagnostic features to alert the user when the electrode needs cleaning Monit...

Page 4: ... 13 Primary display FUNCTIONAL DESCRIPTION 25 Kynar is registered Trademark of Pennwalt Corp Ultem is registered Trademark of General Electric Co HI 1413B Glass body single junction flat tip Viscolene non refillable combination pH electrode Use surface measurement ORP ELECTRODES HI 3131B Glass body refillable combination platinum ORP electrode Use titration HI 3230B Plastic body Ultem gel filled c...

Page 5: ...y r e t t a B s e i r e t t a b e n i l a k l a e z i s A A V 5 1 x 4 e s u s u o u n i t n o c f o s r u o h 0 0 5 x o r p p a f f o o t u A r o s e t u n i m 0 2 e l b a t c e l e s r e s U d e l b a s i d s n o i s n e m i D 4 2 x 1 3 x 7 7 m m 0 6 x 0 8 x 6 9 1 y l n o r e t e m t h g i e W z o 5 1 g 5 2 4 t n e m n o r i v n E F 2 2 1 2 3 C 0 5 0 0 0 1 H R x a m y t n a r r a W s r a e y 2 24...

Page 6: ...pressed to save battery life To disable this feature see Setup Menu section on page 9 pH MEASUREMENTS To take a pH measurement remove the elec trode protective cap and simply submerge the tip 4cm 1 of the electrode and the temperature probe into the sample to be tested If necessary press the RANGE key until the display changes to the pH mode Allow for the electrode to adjust and reading to stabili...

Page 7: ...MEASUREMENTS To perform ORP measurements connect an optional ORP electrode see Accessories section to the meter and turn it ON If necessary enter the mV mode by pressing RANGE until the display changes to mV Submerge the ORP electrode tip 4cm 1 into the sample to be tested and wait a few minutes for the reading to stabilize Measurements within the 699 9 mV range are displayed with 0 1 mV resolutio...

Page 8: ...olution 20 mL sachet 25 pcs HI 70010P pH 10 01 Buffer Solution 20 mL sachet 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL bottle HI 7001M pH 1 68 Buffer Solution 230 mL bottle HI 7004L pH 4 01 Buffer Solution 500 mL bottle HI 7004M pH 4 01 Buffer Solution 230 mL bottle HI 7006L pH 6 86 Buffer Solution 500 mL bottle HI 7006M pH 6 86 Buffer Solution 230 mL bottle HI 7007M pH 7 01 Buffer Solution 50...

Page 9: ...l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C e t y l o r t c e l e h s e r f h t i w l l i f e R y l n o s e d o r t c e l e e l b a l l i f e r s w o h s y a l p s i D e l a c s l l u f g n i k n i l b e u l a v f o t u o g n i d a e R e g n a r s i e l p m a s t a h t k c e h C e g n a r e l b a r u s a e m n i h t i w d n a l e v e l e t y l o r t ...

Page 10: ... to normal measurement mode The LCD will show an empty bargraph to warn the user that the meter is not calibrated 19 For refillable electrodes if the refill solution electrolyte is more than 2 cm 1 below the fill hole add the appropriate Electrolyte Solution MEASUREMENT Rinse the electrode tip with distilled water immerse it 4 cm 1 in the sample and stir gently for a few seconds For a faster respo...

Page 11: ...trument is provided with a real time clock RTC in order to monitor the time elapsed since the last pH calibration The real time clock is reset every time the meter is calibrated and the expired calibration status is triggered when the meter detects a calibration time out The CAL tag will start blinking to warn the user that the meter should be recalibrated The calibration time out can be set see S...

Page 12: ...are cleared the comparison is done with the default values 17 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4 months 120 C Less than 1 month Alkaline Error High concentrations of sodium ions interfere with readings in alkaline solutions The pH at which the interference starts to be significant depends upon the composition of the glass This interference is called alkaline erro...

Page 13: ... 0 0 1 1 5 8 5 8 1 8 7 1 7 1 4 7 8 6 2 0 7 7 8 8 4 7 9 1 9 0 1 0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 16 LCD MESSAGE GUIDE TAGS SYMBOLS Mode tags light up for indicating the corresponding active mode and blink for warning the user SETUP on setup menu mode has been entered CFM blinking ask confirmation of calibration value CAL on ...

Page 14: ...batteries soon The meter is also provided with the BEPS BatteryErrorPreventionSystem feature which automatically turns the instrument off when the battery level is too low to ensure reliable readings At start up the display will show 0 bAtt for a few seconds then the meter automatically turns off Battery replacement must only take place in a non hazardous area using the battery types specified in ...

Reviews: