39
1-MM/DD/YYYY
2-YYYY/MM/DD
3-YYYY-MM-DD
4-Mon DD, YYYY
5-DD-Mon-YYYY
6-YYYY-Mon-DD
-Language name (3 chars, “sfm” when the instrument is in safe mode)
-Checksum (2 chars)
NSLx
Requests the number of logged samples (4 chars).
The command parameter (1 char):
-P-request for pH meter
-T-request for titrator
LODPxxx
Requests the xxxth pH record logged data
LODTxxx
Requests the xxxth titrator record logged data
LODPALL
Requests all the stored records from the pH meter log
LODTALL
Requests all the stored records from the titrator log
The answer string for each record contains:
-Logged mode (2 chars): -00-pH meter (0.1 resolution)
-01-pH meter (0.01 resolution)
-02-titrator
-Measured value status: R-in range, O-over range, U-under range.
-Measured value (7 chars): for the pH meter log this is the measured pH value; for
the titrator log this is the concentration value
-Temperature reading, with two decimal points (7 chars) (only for pH meter log)
-Calibration offset (7 chars) (only for pH meter log)
-Calibration slope (7 chars) (only for pH meter log)
-Acidity unit (1 char) (only for titrator log): 0=%CA, 1=%TA, 2=%MA
-Logged time (12 chars):
yymmddhhmmss
-Temperature probe presence (1 char)
-Checksum (2 chars)
Notes:
-”Err8” is sent if the instrument is not in measurement mode
-”Err6” is sent if the requested range is not available
-”Err4” is sent if the requested set parameter is not available
-”Err3” is sent if the Log on demand is empty
-Invalid commands will be ignored.