26
Verify:
The instrument has been calibrated (pH and pump) before performing a lot of titrations.
An electrode calibration in at least one point (8.20 pH buffer) is recommended.
Sample preparation:
Add to a clean beaker 2 mL fruit
juice precisely measured and fill the beaker up to 50 mL
with distilled water. Place the stir bar into the beaker and
then place the beaker in the appropriate place on the
instrument top.
Notes:
&KHFN WKH S+ RI GLVWLOOHG ZDWHU ,I QHFHVVDU\
adjust the pH to 8.2 by adding 1-2 drops of
titrant solution.
)UXLW MXLFHV VDPSOHV PXVW EH VWRUHG LQ FDSSHG
botlles. Avoid disturbing any sediment in the
sample.
3ODFH WKH HOHFWURGH KROGHU RQ WKH WRS RI WKH EHDNHU
and secure it by turning clockwise.
,PPHDQGWKHWHPSHUDWXUH
probes approximatively
FPµLQWRWKHVDPSOHWREHWHVWHGSD\LQJDWWHQWLRQ
not to touch the stir bar. Use
O-Rings
provided to secure
the pH electrode in holder.
,QVHUW WKH GRVLQJ WLS LQ WKH DSSURSULDWH KROGHU SODFH DQG
immerse it into the sample.
)URP WKH WLWUDWRU PDLQ VFUHHQ SUHVV
Titration
. To enter
titrator main screen from pH meter mode press
Titrator
and then
Titration
,I WKH WXWRULDO LV HQDEOHG WKLV VFUHHQ
will be shown.
3UHVV
Start
to begin the titration process.
TITRATION PROCEDURE
- before adding juice
s
ample.