6
REQUIRED REAGENTS
Code
Description
Quantity/Test
HI 84432 - 50
Titrant
1 mL
HI 84432 - 55
Pump Calibration Solution
2 mL
SPECIFICATIONS
Titrator
Range
Titratable acidity:
g/100 mL
as citric acid : 0.20 - 8.00% CA
g/100 mL
as tartric acid
: 0.23 - 9.30% TA
g/100 mL
as malic acid
: 0.21 - 8.30% MA
Resolution
Titratable acidity:
0.01%
Accuracy
5% of reading ± 0.05
Titration method Acid-base titration
Principle
End point titration : 8.20 pH
Pump debit
0.5 mL/min
Stirring speed
600 rpm
Log data
Up to 50 samples
pH meter
pH meter
-2.0 to 16.0 pH / -2.00 to 16.00 pH
pH Resolution
0.1 pH / 0.01 pH
pH Accuracy
± 0.01 pH
pH Calibration
1, 2 or 3 calibration points;
3 available buffers (4.01; 7.01; 8.20)
Temperature
manual or automatic from
compensation
-20 to 120 °C (-4 to 248 °F)
Log data
Up to 50 samples
Temperature
Range
-20.0 to 120.0 °C (-4.0 to 248.0 °F)
Resolution
0.1 °C
Accuracy
±0.4 °C without probe error
Electrode
HI 1131B (included)
Temperature Probe
HI 7662-M (included)
Environment
0 to 50 °C (32 to 122 °F); max 95% RH non-condensing
Power supply
12 Vdc power adapter
Dimensions
208 × 214 × 163 mm (8.2×8.4×6.4”) (with beaker)
Weight
2200 g (77 oz.)