12
AUTO
mode
The dosage is activated and deactivated ac-
cording with the selected setpoint. The corre-
sponding DOSAGE MODE LED (right side of
the switch) is off.
Be sure that the DOSAGE switch is in AUTO
position when the meter is in normal operat-
ing mode.
ON
mode
The dosage is always activated. The corre-
sponding DOSAGE MODE LED blinks.
INITIAL PREPARATION & INSTALLATION
Material needed:
• a 3-wire power cable to connect HI 8410
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
Ensure the controller has been calibrated. Ensure the D.O. probe is
immersed in the sample to be tested.
The D.O. reading can be displayed in % air saturation or in ppm (mg/L)
according to the range selection switch. The corresponding LED is ON.
For accurate dissolved oxygen measurements a water flow of at least
5-7 cm/sec is required. In this way a constant replenishment of the
oxygen-depleted membrane surface is ensured.
Accurate readings are not possible in still water.
Note:
For some particular applications, such as fish farming, the mem-
brane can be sterilized with stabilized iodine (20 to 50 ppm),
typically used for this purpose.
Note:
For use in harsh environments, it is recommended to protect
the membrane with the optional
HI 76410-0
sleeve. The re-
sponse time will slightly slow down.