
10
•
Recorder Output
(terminals 1, 2, 3)
These contacts are used for connec-
tion to a recorder output. The output
can be 0-20 mA or 4-20 mA and is pro-
portional to the measured Dissolved
Oxygen values.
Connect the “+” wire from the recorder to terminal 1 on the instru-
ment and the other wire (common) to terminal 2 for 4-20 mA re-
corder output or terminal 3 for 0-20 mA recorder output.
Note:
Only one recorder output connection is permited. In order to avoid
malfunction leave unconnected the unused terminal.
REAR CONNECTIONS FOR HI 8410
• Power Connection Terminals
4-screw-terminal-strip for connection
to a 3-wire power cable according to
the indicated voltage (115 or 230V).
• Probe connection
For connection of the
HI76410
DO
probe. Connect the wires according with
the indicated color.
• Set Contacts
Dosing pumps or other control equip-
ment may be connected to the "SET"
(Max. 2A, 240 V) terminals (1, 2).
These contacts act only as a "dry"
switch allowing electrical continuity,
not as a power supply.
CONNECTIONS
CONNECTIONS
CONNECTIONS
CONNECTIONS
CONNECTIONS