43
• Press
CFM
to confirm. The secondary LCD will show “
50.0 ºC
“.•
Submerse the temperature probe (or EC probe) in the second vessel as
near as possible to the calibrated thermometer. Allow a few seconds
for the probe to stabilize.
• Use the
ARROW
keys to set the reading on the secondary LCD to
that of the hot water.
• When the reading is stable and close to the
selected calibration point, the “
CFM
” tag will
blink.
• Press
CFM
to confirm. The instrument will memorize calibration and
restart in measurement mode.
Note: If the reading is not close to the selected calibration point,
“WRONG” tag will blink. Change the temperature probe (or EC
probe) and restart calibration.