EXPIRED CALIBRATION ALARM
Press
CAL
when the calibration time‑out is displayed. Calibration time‑out
(“
OFF
” or “
1
” to “
7
” days) and “
CFM
” will start blinking.
Press the
ARROW
keys to change the calibration time‑out value.
Press
CFM
to save the modified calibration time‑out value.
Press
CAL
to escape without saving.
LOG INTERVAL
Press
CAL
when log interval is displayed. The log interval and “
CFM
”
will start blinking (“
LOG
” for log on demand, “
StAb
” log on stability,
interval in seconds or minutes).
Press the
ARROW
keys to change the custom buffer value.
Press
CFM
to confirm the selection.
Press
CAL
to escape without saving.
FIRST CUSTOM BUFFER
Press
CAL
when
“Cust. Buff. 1”
is displayed. The custom buffer (disabled
–
“no”
or “
0
” to “
16
” pH) and “
CFM
” will start blinking.
Press the
ARROW
keys to change the custom buffer value.
Press
CFM
to save the modified custom buffer value.
Press
CAL
to escape without saving.