28
Up to 700 logged samples can be stored into memory.
200 manually logged records and 500 lot logging records can be stored
in the memory. To select logging type enter SETUP menu.
LOGGING THE CURRENT DATA (manual logging)
Select the manual logging mode in SETUP menu.
To store the current reading into memory press
LOG
while the instrument
is in measurement mode.
The instrument will display the current date on the primary LCD, the
record number on the secondary LCD, “
LOG
” and “
Manual
” tags for a
few seconds (see example below: record No. 11):
followed by the number of free records:
If there are less than 6 memory locations remaining, the record number
and “
Lo
” message will be displayed to alert the user.
If the log space is full (200 records), “
FULL LOG
” message will be
displayed and no more data will be saved.
When
LOG
is pressed, a complete set of information is stored: date, time,
pH, mV, EC, TDS, NaCl temperature and calibration data.
LOGGING FUNCTION