49
• Calibration alarm time out for pH range (2 chars)
• SETUP information (2 chars): 8 bit hexadecimal
encoding.
• 0x01 ‑ beep ON (else OFF)
• 0x04 ‑ degrees Celsius (else degrees Fahrenheit)
• 0x08 ‑ Offset calibration (else Point calibration)
• The number of custom buffers (1 char)
• The custom buffer values, with sign and decimal
point, for each defined custom buffer (7 chars)
• Log type 2 chars
01 ‑ manual log
02 ‑ stability lot log
03 to 14 ‑ the coresponding interval for lot log
(5 s to 180 min)
• cell constant, with sign and decimal point (6 chars)
• TDS factor, with sign and decimal point (5 chars)
• TC coef, with sign and decimal point (5 chars)
• Reference Temperature, with sign and decimal point
(5 chars)
• Temperature Compensation mode (1 char)
NSLx
Requests the number of logged samples.
x = P ‑ request for pH range
M ‑ request for mV range
E ‑ request for EC range
N ‑ request for NaCl range
T ‑ request for TDS range
LODxnnn
: request the”nnn” record of the manual log on the “x” range
LODxALLff
: groups all Log On Demand Records in frames of 8 records
each for the selected range
Command Parameters:
x ‑ range (see Note)
ALL ‑ download all records for the selected range
ff ‑ requested frame number ‑ first frame is labeled 01
LLsxff
: requests information about all lots on the specified range, it sends
the information in frames of 10 lots each (a frame contains information
about 10 lots)
Command Parameters:
x ‑ range (see Note)
ff ‑ requested frame number ‑ first frame is labeled 01