USER MANUAL T04B
A/S Inquiry : 02-3452-8966 / [email protected]
ihanil.com 22
Error 1
Error 2
1) Check the lid closed.
2) Check the Lid lamp status.
1) Remove any heat-generating devices around the centrifuge,
check if the centrifuge vent is blocked or clogged with foreign
substances, and take measures so that the heat generated from
the centrifuge can be dissipated smoothly.
2) Turn off the power, stop using the product for about an hour
(open the lid at this time), and turn the power on again to check.
Motor Overheating : Occurs
when the motor overheats.
Error 3
Error 4
Low Voltage : Occurs when the
supply voltage is below -10%.
1) Check the supply voltage.
2) If it is less than +10% of the rated voltage,
install an AVR to supply the rated voltage.
Error 5
High Voltage : Occurs when the
supply voltage is above +10%.
Error 6
Overspeed : Occurs when 1,000 RPM
or more is higher than the set speed.
Turn off the power and turn it on again to check the operation
status again.
Turn off the power and turn it on again to check the operation
status again.
Firmware Program : Occurs due to
a system error in the control unit.
Error 7
Error 8
Imbalance : Occurs when the
sample balance is not correct.
Error 17
Error 15
Error 9
Turn off the power and turn it on again to check the operation
status again.
Turn off the power and turn it on again to check the operation
status again.
Turn off the power and turn it on again to check the operation
status again.
Lid error : Occurs when the Lid
Sensor does not operate normally.
Error 20 ~ 27
1) Check that the weight of the sample inserted in the rotor is the
same and that it is inserted symmetrically.
2) Check if there is any imbalance in the device due to the level of
the floor, and if there are any factors that cause the device to
move, remove and reinstall to balance.
3) Remove the rotor and wipe off any foreign matter on the shaft
and the connection part, and check that there is no bending of
the motor shaft. If there is no problem, align the threads of the
rotor and the motor shaft and tighten them completely.
4) Check if the tube or bottle is crushed or spilled.
1) Check if the rotor is rotating through the lid center window.
2) Turn the rotor by hand and check the RPM change in the
display window.
RPM Sensing : Occurs when the sensor is
defective or the motor cannot rotate.
Motor Temperature Sensor : Occurs
when the temperature sensor on the
motor is not recognized.
Motor start error: Occurs when the motor
does not reach 200 rpm within 2 seconds
after starting the operation.
Lid Open : Occurs when Lid is
opened during operation.
Check the rotor rotation through the lid's center window.
Communication error: Occurs when
communication between
Main-Display-I/O Boards is not available.
1) Check the supply voltage.
2) If it is more than +10% of the rated voltage,
install an AVR to supply the rated voltage.
6.2 Error message information
If the problem persists after taking the following measures, please contact the service center.
Error code
Description
Action
Summary of Contents for T04B
Page 1: ...T04B LOW SPEED CENTRIFUGE USER MANUAL...
Page 7: ...USER MANUAL T04B A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com 6...
Page 8: ...USER MANUAL T04B 7 ihanil com A S Inquiry 02 3452 8966 techsupport ihanil com...
Page 31: ...A S Inquiry 02 3452 8966 techsupport ihanil com USER MANUAL T04B ihanil com 30 MEMO...
Page 32: ......