4.4 Acceleration/deceleration (ACC/DEC) setting
4.5 Pulse (Short Spin)
Acceleration speed can be set up to 9 steps and deceleration speed up to 10 steps (natural
deceleration:0) for sensitive sample protection and clean layer separation.
1. Press [ACC] to set acceleration of
[DEC] to set deceleration.
2. To enter ACC or DEC value, use the numeric
keypad to enter the setting value and press [Enter].
▶
ACC is set in 1~9 steps
(ACC:9 fastest acceleration section)
▶
DEC is set in 0~9 steps
(DEC:0 natural deceleration)
▶
Press [Enter] → final storage of ACC/DEC values
▶
If you enter the setting value incorrectly, press [←] and re-enter the setting value.
The [Pulse] button is for when press & hold, to accelerate up to maximum speed
and decelerate when the button is not pressed.
1. Set the desired acceleration and deceleration steps.
2. Press the [pulse] button more than 2 sec.
▶
While pressing [Pulse], the equipment will
accelerate up to maximum RPM.
▶
By stop pressing the [Pulse], the equipment
will decelerate to stop.