USER MANUAL T04B
11 ihanil.com
A/S Inquiry : 02-3452-8966 / [email protected]
3.6 Enter rotor ID
Rotor
SO-50-4
AO-50-10
AO-50c-10
SO-100-4
AO-15-24
SO-15-6
SO-MW-a2
Rotor ID
1
2
3
4
5
6
7
Correct tube position
Incorrect tube position
Must enter the rotor ID correctly so that the RPM / RCF can be converted correctly.
1. If you press the Program button for 2 seconds, the Rotor information window appears.
▶
While the Rotor Information window is displayed, enter the Rotor information using the
number buttons.
2. Sample tubes must be placed and mounted symmetrically.
▶
T
he tube filled with the sample, there should be no difference in weight
and the density should be symmetrically and evenly arranged.
▶
Tubes for centrifuges must be used. Max. It should be used after checking the RCF value.
Do not use beyond acceptable standards.