![Hameg HM7044 Manual Download Page 19](http://html1.mh-extra.com/html/hameg/hm7044/hm7044_manual_3419708019.webp)
Commands
19
Subject to change without notice
FUSE
FORMAT F(USE) [(1)(2)(3)(4)] [ON] [OFF]
Set Current-Fuse active
SEL 1,2
>
[
channel 1,2 selected
]
FUSE ON
>
[
channel 1,2 fuse active
]
Set Current-Fuse Number
FUSE 1,2,2,1 >
[
fuse set to 1,2,2,1
]
READ
Format: READ (READOUT/READVALUES)
This command reads all numerical values and state
information of the different channels.
READ
> 00.01V
12.00V
13.22V
14.70V
> 2.787A
0.000A
0.000A
3.000A
> CC -1
CV -2CV F3
OFF F4
Channel I
Channel II
Channel III
Channel IV
OFF:
Channel is switched-off.
CV:
Channel is set into voltage regulation mode
(constant voltage)
CC:
Channel is set into current regulation mode
(constant current)
F/-:
F = Fuse is active / - = Fuse is not active
1,2,3,4:
Channel-Fuse identification number
LOCK
Format: LOCK [ON] [OFF]
Manuel system input lock / unlock
LOCK ON
>
[
keyboard locked
]
LOCK OFF
>
[
keyboard unlocked
]
If the keyboard is locked, the PROGRAMM-LED is flashing.
ON
FORMAT: ON
Switch-on all selected channels (Output on)
OFF
FORMAT: OFF
Switch-off all selected channels (Output off)
ENABLE OUTPUT
FORMAT: EN(ABLE OUTPUT)
see OUTPUT-key
DISABLE OUTPUT
FORMAT: DIS(ABLE OUTPUT)
see OUTPUT-key