background image

34

Subject to change without notice

Trigger coupling

Instrument specific information regarding this item can be
noted in the ”Data Sheet”. The coupling setting (

TRIG. MODE

(20)

) and indication are described under ”Controls and

Readout”.

As the automatic triggering does not work below 20Hz, normal
triggering should be used in DC and LF trigger coupling mode.
The coupling mode and accordingly the frequency range of
the trigger signal should meet the signal requirements.

AC:

     This is the most frequently used trigger mode. The trigger

threshold increases below and above the frequency limits
mentioned in the data sheet. This filter cuts off both the
DC content of the trigger signal and the lowest frequency
range.

DC:

     In this coupling mode the trigger signal is coupled

galvanically to the trigger unit if normal triggering (NM) is
present. Therefore there is no low frequency limit.

DC triggering is recommended if the signal is to be
triggered with quite slow processes or if pulse signals
with constantly changing pulse duty factors have to be
displayed.

HF:

     In this coupling mode the transmission range equals a

high pass filter. It cuts off the DC content of the trigger
signal and the lower frequency range.

LF:

      LF trigger coupling has a low pass filter function

characteristic. As in DC trigger coupling, there is no limit
for the pass frequency range in connection with normal
triggering.

The LF trigger coupling is often more suitable for low
frequency signals than DC trigger coupling because the
noise components of the trigger signals are strongly
suppressed. This avoids or reduces, under borderline
conditions, jitter or double traces especially with very low
signal voltages. The trigger threshold rises continuously
above the pass band.

Tv-L:

The built in active TV Sync Separator provides the

separation of line sync pulses from the video signal.

Even distorted video signals are triggered and displayed
in a stable manner. This mode is described under paragraph
”Triggering of video signals”.

Tv-F:

The built in active TV Sync Separator also provides the

separation of frame sync pulses from the video signal.
Even distorted video signals are triggered and displayed
in a stable manner.

This mode is described under paragraph ”Triggering of
video signals”.

~

:

     Indicates “line/mains triggering” and is described under

the paragraph of the same name.

Triggering of video signals

In     

Tv-L

     and     

Tv-F

     trigger coupling mode the instrument is

automatically set to automatic triggering and the trigger point
indicator is switched off. As only the separated synchronization
pulses are used for triggering the relationship between the
displayed signal and the trigger signal is lost. In     

TV-F

     mode

interference may occur if chopped     

DUAL

     mode is chosen or

the readout is active.

The automatic triggering works above 20Hz. The failure of
automatic triggering at frequencies below 20Hz is abrupt.
However, it is not signified by the trigger indicator LED which
may still be blinking. Break down of triggering is best
recognizable at the left screen edge (the start of the trace in
differing display height).

The automatic peak (value) triggering operates over all
variations or fluctuations of the test signal above 20Hz.
However, if the pulse duty factor of a square wave signal
exceeds a ratio of 100:1, switching over to normal triggering
will be necessary. Automatic triggering is practicable with
internal and external trigger voltage.

Normal Triggering

Information specific to the instrument is given in the sections

NM - AT -  (9)

,

 LEVEL (11

) and     

TRIG. MODE (20)

     in the

paragraphs ”Controls and Readout”. The time fine adjustment
(((((VAR.), and the hold off time setting assist in triggering under
specially difficult signal conditions.

With normal triggering, the sweep can be started by AC

 AC

 AC

 AC

 AC

signals within the frequency range defined by the trigger
coupling setting.

In the absence of an adequate trigger signal or when the trigger
controls (particularly the trigger     

LEVEL

     control) are misadjusted,

no trace is visible.

When using the internal normal triggering mode, it is possible
to trigger at any amplitude point of a signal edge, even with
very complex signal shapes, by adjusting the trigger     

LEVEL

control. If the signal applied at the Y input is used for triggering
(internal trigger source), its adjusting range is directly
dependent on the display height, which should be at least
0.5div. If it is smaller than 1div, the trigger     

LEVEL

     adjustment

needs to be operated with a sensitive touch. In the external
normal triggering mode, the same applies to approx. 0.3Vpp
external trigger voltage amplitude.

Other measures for triggering of very complex signals are the
use of the time base variable control and     

HOLD OFF

     time

control, mentioned below.

/ \

 

SLOPE

Please note item (9) in section ”Controls and Readout”
for instrument specific information.

The actual slope setting is displayed in the readout. The
setting is not changed by the

 

AUTOSET 

AUTOSET 

AUTOSET 

AUTOSET 

AUTOSET function. The slope

setting can be changed for the delay timebase trigger unit
in delay mode if the delay trigger function is active. The
previous slope setting for the undelayed time base trigger
is stored and still active. For further information please note
”Controls and Readout”.

The time base generator can be triggered by a rising or falling
edge of the test signal. Whether the rising or the falling edge
is used for triggering, depends on the slope direction setting.
This is valid with automatic and normal triggering. The positive
slope direction means an edge going from a negative potential
and rising to a positive potential. This has nothing to do with
zero or ground potential or absolute voltage values. The positive
slope may also lie in a negative part of a signal.

However the trigger point may be varied within certain limits
on the chosen edge using the LEVEL 

 LEVEL 

 LEVEL 

 LEVEL 

 LEVEL control. The slope

direction is always related to the input signal and the non
inverted display.

Triggering and time base

Summary of Contents for HM507

Page 1: ...Instruments Oscilloscope HM507 E N G L I S H HANDBUCH MANUAL MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...tomatic Peak value Triggering 33 Normal Triggering 34 Slope 34 Trigger coupling 34 Triggering of video signals 34 Line Mains triggering 35 Alternate triggering 35 External triggering 35 Trigger indica...

Page 4: ...place via the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of t...

Page 5: ...Interface 50MHz Analog Digital Oscilloscope HM507 Oscilloscope Specifications Vertical Deflection analog digital Operating modes Channel I or CH II separate CH I and II alternate or chopped 0 5MHz Sum...

Page 6: ...be negated by the use of an extension cord without a protective conductor The mains line plug should be inserted before connections are made to measuring circuits The grounded accessible metal parts c...

Page 7: ...data are based Purchase of the new HAMEG scope tester HZ 60 which despite its low price is highly suitable for tasks of this type is very much recommended The exterior of the oscilloscope should be cl...

Page 8: ...lue is 11 higher The gain reduction error can not be defined exactly as the 3dB bandwidth of the amplifiers differ between 40MHz and 42MHz For sinewave signals the 6dB limit is approx 50MHz When exami...

Page 9: ...se four values are not freely selectable They have to be within the following limits L between 0 2 and 10div if possible 4 to 10div T between 10ns and 5s F between 0 5Hz and 100MHz Tc between 100ns di...

Page 10: ...ng the probe should always be used The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier Because of their separate manufacture all attenuato...

Page 11: ...at 1 Component Tester is switched off 2 The following settings are present under MAIN MENU SETUP INFO MISCELLANEOUS 2 1 CONTROL BEEP and ERROR BEEP activated x 2 2 QUICK START not activated 3 The scre...

Page 12: ...memories These can be used by the operator to save instrument settings and to recall them SAVE Press and hold the RECALL SAVE button to start a storage process This causes the SAVE menu Standard menu...

Page 13: ...e determined by taking the deflection coefficient the probe factor and the trace position change with respect to the previous 0 Volt position into account 0 Volt Symbol The READOUT indicates the 0 Vol...

Page 14: ...tomatic or automatic peak value triggering is selected AT Automatic triggering can be carried out with or without peak capture In both cases the LEVEL control 11 is effective and the trace is visible...

Page 15: ...direction ccw The available range is from 1mV div up to 20V div The deflection coefficients and additional information regarding the active channel s are displayed in the readout e g Y1 deflection coe...

Page 16: ...coefficient display as the time base is switched off 2 Controls and readout display switched off for trigger source slope and coupling 3 Y POS CURS I knob inactive for X shift use X POS 12 knob 4 X MA...

Page 17: ...nel II and the READOUT displays Y2 trigger slope trigger coupling The last function setting of the VOLTS DIV 18 knob remains unchanged All channel II related controls are active if the input 28 is not...

Page 18: ...zero and maximum can be set The signal displayed to the left of the trigger point symbol shows the pre history the right side displays the run of the curve after the trigger event POST TRIGGER As defi...

Page 19: ...cient and can be selected by the TIME DIV knob It is recommended to capture and display the signals in DUAL mode and choose a suitable time coefficient which enables you to see the higher frequency si...

Page 20: ...7 divisions The blanked section serves as a guide to the delay time The delay time is based on the current time deflection coefficient setting and can also be coarsely set with the TIME DIV control ra...

Page 21: ...in conjunction with an operating mode instead of a measuring value the READOUT indicates n a not applicable For example the READOUT displays t n a if t measurement is selected in combination with XY m...

Page 22: ...er and trigger amplifier frequency responses there are also deviations with respect to the signal display If relatively low frequency signals 20Hz are present the measurement value continuously change...

Page 23: ...maximum symbol positioning and measuring accuracy first set the signal slope to the screen center X POS 12 and then activate X magnifier X MAG x10 13 To avoid CURSOR line and symbol changes after each...

Page 24: ...reference value The measured value is calculated from the distance between the short CURSOR lines compared to the reference value This method is suitable to determine e g the oscilloscope s frequency...

Page 25: ...as Y1 channel I Y2 channel II and M MATH signal their availability depends on the actual operating mode 1 If DUAL or XY mode is present in combination with CURSOR voltage measurement CURSOR MEASURE V...

Page 26: ...m highlighted the equation to be displayed later can be chosen MAT1 MAT2 MAT3 and the mathematic function can be switched ON x or OFF by pressing the x pushbutton CAL SEL Pressing and holding opens th...

Page 27: ...ith double function STOR ON Pressing and holding switches over from analog to storage digital mode and vice versa In the case of CT Component Tester mode active this mode must be left before it is pos...

Page 28: ...right by the readout The following description assumes that the trigger conditions are met in Refresh rfr and it s submodes Envelope env and Average avm 41 1 1 rfr Refresh mode In this mode as in anal...

Page 29: ...ease note the HO79 6 manual E MAIN MENU The instrument software contains several menus The following menus submenus and menu items are available 1 ADJUSTMENT contains 1 1 AUTO ADJUSTMENT with the item...

Page 30: ...ode is not selected XY not displayed in the readout To obtain the maximum life from the cathode ray tube the minimum intensity setting necessary for the measurement in hand and the ambient light condi...

Page 31: ...n should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not diffi...

Page 32: ...ue to phase shift do not use too high a test frequency It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the t...

Page 33: ...r external trigger voltage A base line will always be displayed even with no signal With an applied AC signal peak value triggering enables the user to select the trigger point on the displayed signal...

Page 34: ...er it is not signified by the trigger indicator LED which may still be blinking Break down of triggering is best recognizable at the left screen edge the start of the trace in differing display height...

Page 35: ...frequency of the Y signal and is recommended for all mains line synchronous signals This also applies within certain limits to whole number multiples or fractions of the line frequency Line triggering...

Page 36: ...is possible with certain pulse signals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use o...

Page 37: ...so possible in the dual channel addition and difference modes In chopped DUAL mode if after switching over to del or dTr the time deflection coefficient is reduced TIME DIV the channel switching mode...

Page 38: ...ntrols and Readout The instrument has a built in electronic Component Tester which is used for instant display of a test pattern to indicate whether or not components are faulty It can be used for qui...

Page 39: ...relatively small capacitance or a relatively high inductance A vertical ellipse indicates a low impedance or a relatively large capacitance or a relatively small inductance A sloping ellipse means tha...

Page 40: ...e displayed as a complete curve Narrow pulses with low repetition rates do not cause intensity reduction Documentation and processing of captured signals is easily possible In comparison with analog m...

Page 41: ...ired graticule position The trigger point symbol then should be set above or below the 0 Volt position line according to the expected voltage of the event to be captured Whether the slope selection is...

Page 42: ...ting the sampling rate is too low the display of an alias signal may occur As described under Controls andReadout item 22 TIME DIV 22 3 AL thereadoutdisplays a warning if less than 2 samples are taken...

Page 43: ...ching the oscilloscope off or transmitting the command RM 0 from the PC to the oscilloscope or depressing the AUTOSET LOCAL pushbutton if in unlocked condition command LK 1 was not sent After the remo...

Page 44: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: