background image

10

Subject to change without notice

with approximately constant group delay (therefore good pulse
transmission performance) the following numerical relationship
between rise time tr (in ns) and bandwidth B (in MHz) applies:

Connection of Test Signal

In most cases briefly depressing the AUTO SET causes a useful
signal related instrument setting. The following explanations
refer to special applications and/or signals, demanding a manual
instrument setting. The description of the controls is explained
in the section “controls and readout”.

Caution:
When connecting unknown signals to the oscilloscope
input, always use a x10 probe, automatic triggering
and set the input coupling switch to DC (readout). The
attenuator should initially be set to 20V/div.

Sometimes the trace will disappear after an input signal has
been applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160Vpp and
the deflection coefficient (

VOLTS/DIV.

) in calibrated condition,

an attenuator probe must be inserted before the vertical input.
If, after applying the signal, the trace is nearly blanked, the
period of the signal is probably substantially longer than the set
time deflection coefficient (

TIME/DIV.

). It should be switched

to an adequately larger time coefficient.

The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as 

HZ32

or

 HZ34

, or reduced through a x10 or x100 attenuator probe. The

use of test cables with high impedance circuits is only
recommended for relatively low frequencies (up to approx.
50kHz). For higher frequencies, the signal source must be of low
impedance, i.e. matched to the characteristic resistance of the
cable (as a rule 50

). Especially when transmitting square and

pulse signals, a resistor equal to the characteristic impedance
of the cable must also be connected across the cable directly
at the Y-input of the oscilloscope. When using a 50

 cable such

as the

 HZ34

, a 50

 through termination type 

HZ22 

is available

from

 HAMEG

. When transmitting square signals with short rise

times, transient phenomena on the edges and top of the signal
may become visible if the correct termination is not used. A
terminating resistance is sometimes recommended with sine
signals as well. Certain amplifiers, generators or their attenuators
maintain the nominal output voltage independent of frequency
only if their connection cable is terminated with the prescribed
resistance. Here it must be noted that the terminating resistor

HZ22 

will only dissipate a maximum of 2Watts. This power is

reached with 10Vrms or  at 28.3Vpp with sine signal. If a x10 or
x100 attenuator probe is used, no termination is necessary. In
this case, the connecting cable is matched directly to the high
impedance input of the oscilloscope. When using attenuators
probes, even high internal impedance sources are only slightly
loaded (approx. 10M

 II 12pF or 100M

 II 5pF with

 HZ53

).

Therefore, if the voltage loss due to the attenuation of the probe
can be compensated by a higher amplitude setting, the probe
should always be used. The series impedance of the probe
provides a certain amount of protection for the input of the
vertical amplifier. Because of their separate manufacture, all
attenuator probes are only partially compensated, therefore
accurate compensation must be performed on the oscilloscope
(see Probe compensation ).

Standard attenuator probes on the oscilloscope normally reduce
its bandwidth and increase the rise time. In all cases where the
oscilloscope bandwidth must be fully utilized (e.g. for pulses

pressed X-MAG. (x10) button: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10

-6

) = 25MHz,

required period T = 1:(25x10

6

) = 40ns

.

If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(X-MAG. (x10) active). In this case, the time interval of interest can
be shifted to the screen center using the X-POS. control.

When investigating pulse or square waveforms, the critical
feature is the risetime of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly influence
the measuring accuracy, the risetime is generally measured
between 10% and 90% of the vertical pulse height. For
measurement, adjust the Y deflection coefficient using its varia-
ble function (uncalibrated) together with the Y-POS. control so
that the pulse height is precisely aligned with the 0% and 100%
lines of the internal graticule. The 10% and 90% points of the
signal will now coincide with the 10% and 90% graticule lines.
The risetime is given by the product of the horizontal distance in
div between these two coincident points and the calibrated time
coefficient setting. The fall time of a pulse can also be measured
by using this method.

The following figure shows correct positioning of the oscilloscope
trace for accurate risetime measurement.

With a time coefficient of 10ns/div (X x10 magnification active),
the example shown in the above figure results in a total measured
risetime of

t

tot

 = 1.6div x 10ns/div = 16ns

When very fast risetimes are being measured, the risetimes of
the oscilloscope amplifier and of the attenuator probe has to be
deducted from the measured time value. The risetime of the
signal can be calculated using the following formula.

In this t

tot

 is the total measured risetime, t

osc

 is the risetime of the

oscilloscope amplifier (approx. 8.75ns), and t

p

 the risetime of the

probe (e.g. = 2ns). If t

tot

 is greater than 100ns, then t

tot

 can be taken

as the risetime of the pulse, and calculation is unnecessary.

Calculation of the example in the figure above results in a
signal risetime

t

r

 = 

16

2

 - 8.75

2

 - 2

= 13.25ns

The measurement of the rise or fall time is not limited to the trace
dimensions shown in the above diagram. It is only particularly simple
in this way. In principle it is possible to measure in any display position
and at any signal amplitude. It is only important that the full height
of the signal edge of interest is visible in its full length at not too great
steepness and that the horizontal distance at 10% and 90% of the
amplitude is measured. If the edge shows rounding or overshooting,
the 100% should not be related to the peak values but to the mean
pulse heights. Breaks or peaks (glitches) next to the edge are also
not taken into account. With very severe transient distortions, the
rise and fall time measurement has little meaning. For amplifiers

Type of signal voltage

Summary of Contents for HM507

Page 1: ...Instruments Oscilloscope HM507 E N G L I S H HANDBUCH MANUAL MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...tomatic Peak value Triggering 33 Normal Triggering 34 Slope 34 Trigger coupling 34 Triggering of video signals 34 Line Mains triggering 35 Alternate triggering 35 External triggering 35 Trigger indica...

Page 4: ...place via the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of t...

Page 5: ...Interface 50MHz Analog Digital Oscilloscope HM507 Oscilloscope Specifications Vertical Deflection analog digital Operating modes Channel I or CH II separate CH I and II alternate or chopped 0 5MHz Sum...

Page 6: ...be negated by the use of an extension cord without a protective conductor The mains line plug should be inserted before connections are made to measuring circuits The grounded accessible metal parts c...

Page 7: ...data are based Purchase of the new HAMEG scope tester HZ 60 which despite its low price is highly suitable for tasks of this type is very much recommended The exterior of the oscilloscope should be cl...

Page 8: ...lue is 11 higher The gain reduction error can not be defined exactly as the 3dB bandwidth of the amplifiers differ between 40MHz and 42MHz For sinewave signals the 6dB limit is approx 50MHz When exami...

Page 9: ...se four values are not freely selectable They have to be within the following limits L between 0 2 and 10div if possible 4 to 10div T between 10ns and 5s F between 0 5Hz and 100MHz Tc between 100ns di...

Page 10: ...ng the probe should always be used The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier Because of their separate manufacture all attenuato...

Page 11: ...at 1 Component Tester is switched off 2 The following settings are present under MAIN MENU SETUP INFO MISCELLANEOUS 2 1 CONTROL BEEP and ERROR BEEP activated x 2 2 QUICK START not activated 3 The scre...

Page 12: ...memories These can be used by the operator to save instrument settings and to recall them SAVE Press and hold the RECALL SAVE button to start a storage process This causes the SAVE menu Standard menu...

Page 13: ...e determined by taking the deflection coefficient the probe factor and the trace position change with respect to the previous 0 Volt position into account 0 Volt Symbol The READOUT indicates the 0 Vol...

Page 14: ...tomatic or automatic peak value triggering is selected AT Automatic triggering can be carried out with or without peak capture In both cases the LEVEL control 11 is effective and the trace is visible...

Page 15: ...direction ccw The available range is from 1mV div up to 20V div The deflection coefficients and additional information regarding the active channel s are displayed in the readout e g Y1 deflection coe...

Page 16: ...coefficient display as the time base is switched off 2 Controls and readout display switched off for trigger source slope and coupling 3 Y POS CURS I knob inactive for X shift use X POS 12 knob 4 X MA...

Page 17: ...nel II and the READOUT displays Y2 trigger slope trigger coupling The last function setting of the VOLTS DIV 18 knob remains unchanged All channel II related controls are active if the input 28 is not...

Page 18: ...zero and maximum can be set The signal displayed to the left of the trigger point symbol shows the pre history the right side displays the run of the curve after the trigger event POST TRIGGER As defi...

Page 19: ...cient and can be selected by the TIME DIV knob It is recommended to capture and display the signals in DUAL mode and choose a suitable time coefficient which enables you to see the higher frequency si...

Page 20: ...7 divisions The blanked section serves as a guide to the delay time The delay time is based on the current time deflection coefficient setting and can also be coarsely set with the TIME DIV control ra...

Page 21: ...in conjunction with an operating mode instead of a measuring value the READOUT indicates n a not applicable For example the READOUT displays t n a if t measurement is selected in combination with XY m...

Page 22: ...er and trigger amplifier frequency responses there are also deviations with respect to the signal display If relatively low frequency signals 20Hz are present the measurement value continuously change...

Page 23: ...maximum symbol positioning and measuring accuracy first set the signal slope to the screen center X POS 12 and then activate X magnifier X MAG x10 13 To avoid CURSOR line and symbol changes after each...

Page 24: ...reference value The measured value is calculated from the distance between the short CURSOR lines compared to the reference value This method is suitable to determine e g the oscilloscope s frequency...

Page 25: ...as Y1 channel I Y2 channel II and M MATH signal their availability depends on the actual operating mode 1 If DUAL or XY mode is present in combination with CURSOR voltage measurement CURSOR MEASURE V...

Page 26: ...m highlighted the equation to be displayed later can be chosen MAT1 MAT2 MAT3 and the mathematic function can be switched ON x or OFF by pressing the x pushbutton CAL SEL Pressing and holding opens th...

Page 27: ...ith double function STOR ON Pressing and holding switches over from analog to storage digital mode and vice versa In the case of CT Component Tester mode active this mode must be left before it is pos...

Page 28: ...right by the readout The following description assumes that the trigger conditions are met in Refresh rfr and it s submodes Envelope env and Average avm 41 1 1 rfr Refresh mode In this mode as in anal...

Page 29: ...ease note the HO79 6 manual E MAIN MENU The instrument software contains several menus The following menus submenus and menu items are available 1 ADJUSTMENT contains 1 1 AUTO ADJUSTMENT with the item...

Page 30: ...ode is not selected XY not displayed in the readout To obtain the maximum life from the cathode ray tube the minimum intensity setting necessary for the measurement in hand and the ambient light condi...

Page 31: ...n should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not diffi...

Page 32: ...ue to phase shift do not use too high a test frequency It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the t...

Page 33: ...r external trigger voltage A base line will always be displayed even with no signal With an applied AC signal peak value triggering enables the user to select the trigger point on the displayed signal...

Page 34: ...er it is not signified by the trigger indicator LED which may still be blinking Break down of triggering is best recognizable at the left screen edge the start of the trace in differing display height...

Page 35: ...frequency of the Y signal and is recommended for all mains line synchronous signals This also applies within certain limits to whole number multiples or fractions of the line frequency Line triggering...

Page 36: ...is possible with certain pulse signals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use o...

Page 37: ...so possible in the dual channel addition and difference modes In chopped DUAL mode if after switching over to del or dTr the time deflection coefficient is reduced TIME DIV the channel switching mode...

Page 38: ...ntrols and Readout The instrument has a built in electronic Component Tester which is used for instant display of a test pattern to indicate whether or not components are faulty It can be used for qui...

Page 39: ...relatively small capacitance or a relatively high inductance A vertical ellipse indicates a low impedance or a relatively large capacitance or a relatively small inductance A sloping ellipse means tha...

Page 40: ...e displayed as a complete curve Narrow pulses with low repetition rates do not cause intensity reduction Documentation and processing of captured signals is easily possible In comparison with analog m...

Page 41: ...ired graticule position The trigger point symbol then should be set above or below the 0 Volt position line according to the expected voltage of the event to be captured Whether the slope selection is...

Page 42: ...ting the sampling rate is too low the display of an alias signal may occur As described under Controls andReadout item 22 TIME DIV 22 3 AL thereadoutdisplays a warning if less than 2 samples are taken...

Page 43: ...ching the oscilloscope off or transmitting the command RM 0 from the PC to the oscilloscope or depressing the AUTOSET LOCAL pushbutton if in unlocked condition command LK 1 was not sent After the remo...

Page 44: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: