P2010:
Variable balance CH II
– Press CH I/II pushbutton (IN position) to select channel II.
– Locate and identify P2010 on YP-Board.
– Set input attenuator CH II to 5 mV/div.
– Press channel II GND pushbutton (IN position).
– Set trace with Y-POS II control to the horizontal centre line
of the graticule.
– Rotate variable gain knob of channel II continuously fully
counter clockwise (left stop) and clockwise (right stop) and
adjust P2010 for no trace movement.
CW incorrect
CCW incorrect
CW correct
CCW correct
22
P2009:
Invert balance CH II
– Press CH I/II pushbutton (IN position) to select channel II.
– Locate and identify P2009 on YP-Board.
– Set input attenuator CH II to 5 mV/div.
– Press channel II GND pushbutton (IN position).
– Set trace with Y-POS II control to the horizontal centre line
of the graticule.
– Press and release the INV. pushbutton and adjust P2009
for no trace movement.
Not inverted incorrect
Inverted incorrect
Not inverted correct
Inverted correct