1) RV1001: Adjustment of +146 Volt supply
2) +175 Volt check
3) RV1003: Adjustment of +12 Volt supply
4) -6 Volt accuracy check
5) Y plate voltage check
6) R6013: CRT minimum intensity
7) RV1005: Focus symmetry
8) R6024: Astigmatism correction
9) RV4002: Focus control
10) P2007: Y-Gain CH I at 5 mV/div
11) P2002: Y-Gain CH I at 2 mV/div
12) P2003: Channel I Y-MAG. x5 balance
13) P2001: 100 Hz square wave CH I
14) P2005: Variable balance CH I
15) C2011/C2007: 10:1 Attenuator compensation CH I
16) C2009/C2010: 100:1 Attenuator compensation CH I
17) P2011: Y-Gain CH II at 5 mV/div
18) P2015: Y-Gain CH II at 2 mV/div
19) P2016: Channel II Y-MAG. x5 balance
20) P2014: 100 Hz square wave CH II
21) P2010: Variable balance CH II
22) P2009: Invert balance CH II
23) C2110/C2107: 10:1 Attenuator compensation CH II
24) C2108/C2103: 100:1 Attenuator compensation CH II
25) P2012: ADDition balance
26) RV4005: Sawtooth start position
27) RV9004: X-Magnifi cation x1
28) CV4001: 0.5 μs/div time base adjustment
29) RV4004: 50 μs/div time base adjustment
30) RV4003: 50 ms/div time base adjustment
31) RV9003: X-Magnifi cation x10
32) C9026, C9027: X-Magnifi cation x10 linearity
33) RV9002: X-Gain Adjustment in XY-Mode
34) RV4001: Trigger-Symmetry
35) P2006: DC-Triggering CH I
36) P2008: DC-Triggering CH II
37) CV9001/9002, RV9006/9005: Y-Final Amplifi er
38) Y-Amplifi er Bandwidth Check
39) RV9001: Component-Tester Y-Offset
40) VR7003: Trace Rotation check
41) RV4501: Component-Tester trace inclination
42) RV4502: Calibrator Output
43) XY mode: X-Amplifi er Bandwidth Check
44) Trigger Filter Check
45) Trigger Bandwidth Check
46) External Trigger Check
47) Video Trigger Check
48) Invert-Balance Check CH II
49) Variable-Balance Check CH I
50) Variable-Balance Check CH II
NOTE
The adjustment procedures assume that the instrument had
once been properly adjusted in the factory and adjustments are
required due to component aging or the replacement of defective
components.
Before starting each adjustment procedure, set the oscilloscope
to the following basic settings:
– POWER pushbutton pressed (IN position).
– All other pushbuttons (Out Position!).
– Rotate the three variable controls (TIME/DIV. and VOLTS/
DIV.) to their (calibrated) detent positions.
– Set TIME/DIV. switch to 0.2 ms/div.
– Set both VOLTS/DIV switches to 5 mV/div.
– Rotate the HOLD OFF knob fully counter clockwise.
– Set trigger coupling lever to AC.
– Set all other controls to their midrange positions.
If different settings are required, it is specifi ed in each adjust-
ment.
A d j u s t m e n t P r o c e d e r e H M 3 0 3 - 6