38
Subject to change without notice
– Check that trigger starts with fi eld sync. pulses; approx.
2.5 frames should be visible.
– Change polarity of the video signal.
– Press SLOPE pushbutton (IN position).
– Again approx. 2.5 frames should be displayed triggered.
– Set time base to 20 μs/div for TV line triggering.
– Now approx. 3 lines should be visible.
– Change polarity of video signal.
– Release SLOPE pushbutton to the OUT position.
– Again 3 lines should be displayed triggered.
Note:
The switch over from frame to line TV triggering is perfor-
med by the time base switch. In time base settings from
0.2 s/div to 0.2 ms/div TV frame triggering is selected. Line
triggering is present from 0.1 ms/div to 0.1 μs/div.
35. Line Trigger
– Set time base to 5 ms/div.
– Connect 10:1 probe to CH I input.
– Place the probe tip close to the oscilloscope mains cable,
so that it can pick up the more or less distorted line/mains
signal.
– Change attenuator CH I setting for a suitable display height
of at least 1 div.
– Press AT/NM and ALT pushbuttons (IN position).
Note:
This combination is marked with a sine symbol indicating
line/mains triggering.
– Check that the Trigger-LED (TR) is ON and the line/mains
signal is stable triggered.
– Disconnect 10:1 probe from CH I input.
– Check that the Trigger-LED (TR) is still ON although no
signal is applied.
– Release AT/NM and ALT pushbuttons (OUT position).
– Check that the Trigger-LED (TR) is OFF.
36. Trigger Filter Check
– Set time base to 1 ms/div.
– Connect a 1 kHz sine wave signal, 40 mV
pp
amplitude,
to input CH I and check for full screen defl ection (80 mm
at 5 mV/div).
– Set input attenuator CH I to 50 mV/div and check for 8 mm
display height.
– Set trigger coupling to AC, DC and LF. The signal must
always trigger.
– Set time base to 50 μs/div.
– Set sine wave generator to 50 kHz and 40 mV
pp
output
amplitude and check for 8mm display height.
– Set trigger coupling to AC and DC. The signal must always
trigger.
– Select LF trigger coupling. Now the signal should not trig-
ger.
37. Component-Tester
– Press Comp. Tester pushbutton (IN position).
– Check that a horizontal trace of approximately 8 div. length
is displayed.
– Connect a short piece of wire between both “COMP. TES-
TER” sockets.
– Check that the trace is displayed in vertical direction with
approximately 6 div. height.
38. Calibrator Output
– Connect 10:1 probe to CH I input.
– Check that attenuator CH I is set to 5 mV/div.
– Insert the probe tip into the 0.2 V
pp
calibrator signal output
socket.
– Check that the 1 kHz / 1 MHz pushbutton is released (OUT
position).
– Set time base to 0.2 ms/div.
– Check that approximately 2 signal periods are visible.
– Press 1 kHz / 1 MHz pushbutton (IN position).
– Set time base to 0.2 μs/div.
– Check that once again approximately 2 signal periods are
visible.
Spare-Part List HM303-6
P e r f o r m a n c e C h e c k H M 3 0 3 - 6
Spare Part
Spare Part Number
CRT Board
29-1000-0004
CRT Modul (Tube)
29-1000-0008
FC Board
29-1000-0001
TB Board
29-1000-0007
PS Board
29-1000-0002
AT Board
29-1000-0003
CC Board
29-1000-0005
XY Board
29-1000-0006
Front Cover
29-1000-0011
Rear Cover
29-1000-0010
Casing with Handle
29-1000-0009
Knob Set
29-1000-0013
Screenfi lterpane
29-1000-0012