46
Subject to change without notice
13
P2001:
100 Hz square wave CH I
– Locate and identify P2001 on YP-Board.
– Connect a 25 mV
pp
/ 100 Hz square wave signal via a 50 Ω
cable and a 50 Ω through-terminator to input channel I.
– Set input attenuator CH I to 5 mV/div.
– Set input coupling to DC.
– Set time base to 2 ms/div.
– Adjust P2001 for a fl at top on the square wave signal.
Correct
Overcompensate
Undercompensate
14
P2005:
Variable balance CH I
– Locate and identify P2005 on YP-Board.
– Set input attenuator CH I to 5 mV/div.
– Press channel I GND pushbutton (IN position).
– Set trace with Y-POS I control to the horizontal centre line
of the graticule.
– Rotate variable gain knob of channel I continuously fully
counter clockwise (left stop) and clockwise (right stop)
and adjust P2005 for equal trace position (no vertical trace
shift).
CW incorrect
CCW incorrect
CW correct
CCW correct