Hameg HM303-6 Manual Download Page 14

14

Subject to change without notice

Phase difference measurement in DUAL mode

t

= horizontal spacing of the zero transitions in div.

T

= horizontal spacing for one period in div.

In the example illustrated, 

t

 = 3div. and 

T

 = 10div. The phase

difference in degrees is calculated from

or expressed in radians

Relatively small phase angles at not too high frequencies can be
measured more accurately in the X-Y mode with Lissajous figures.

Measurement of an amplitude modulation

The momentary amplitude u at time t of a HF-carrier voltage, which
is amplitude modulated without distortion by a sinusoidal AF
voltage, is in accordance with the equation

where

UT

= unmodulated carrier amplitude

2

πππππ

= angular carrier frequency

ω

ω

ω

ω

ω

2

πππππ

= modulation angular frequency

m

= modulation factor (i.a. 

 1  100%).

The lower side frequency 

F

-

f

  and the upper side frequency 

F+f

  arise

because of the modulation apart from the carrier frequency 

F

.

Figure 1.

Amplitude and frequency spectrum for AM display (m = 50%)

The display of the amplitude-modulated HF oscillation can be
evaluated with the oscilloscope provided the frequency spectrum
is inside the oscilloscope bandwidth. The time base is set so that
several cycles of the modulation frequency are visible. Strictly
speaking, triggering should be external with modulation frequency
(from the AF generator or a demodulator). However, internal
triggering is frequently possible with normal triggering (

AT/NM

button depressed) using a suitable 

LEVEL

 setting and possibly

also using the time variable adjustment.

Oscilloscope setting for a signal according to figure 2:

Depress no buttons.

Y: CH I; 20mV/div.; AC

.

TIME/DIV.: 

0.2ms/div.

Triggering: 

NM 

(NORMAL);

with 

LEVEL

-setting;

internal (or external) triggering.

Figure 2.
Amplitude modulated oscillation:

F

 = 1 MHz; 

f

 = 1 kHz;

m

 = 50 %; 

UT

 = 28.3 mV

rms

.

If the two values 

a

 and 

b

 are read from the screen, the

modulation factor is calculated from

where

a = U

T

 (1+m)

 and 

b = U

T

 (1-m).

The variable controls for amplitude and time can be set arbitrarily
in the modulation factor measurement. Their position does not
influence the result.

Triggering and time base

Time related amplitude changes on a measuring signal (AC
voltage) are displayable in Yt-mode. In this mode the signal
voltage deflects the beam in vertical direction while the time
base generator moves the beam from the left to the right of the
screen (time deflection).

Normally there are periodically repeating waveforms to be
displayed. Therefore the time base must repeat the time
deflection periodically too. To produce a stationary display, the
time base must only be triggered if the signal height and slope
condition coincide with the former time base start conditions.
A DC voltage signal can not be triggered as it is a constant signal
with no slope.

Triggering can be performed by the measuring signal itself
(internal triggering) or by an external supplied but synchronous
voltage (external triggering).

The trigger voltage should have a certain minimum amplitude.
This value is called the trigger threshold. It is measured with a
sine signal. When the trigger voltage is taken internally from the
test signal, the trigger threshold can be stated as vertical
display height in div., through which the time base generator
starts, the display is stable, and the trigger LED lights.

The internal trigger threshold of the 

HM303-6

 is given as £

0.5div. When the trigger voltage is externally supplied, it can be
measured in V

pp

 at the 

TRIG. EXT.

 socket. Normally, the trigger

threshold may be exceeded up to a maximum factor of 20.

The 

HM303-6

 has two trigger modes, which are characterized

in the following.

Operating modes of the vertical amplifiers

Triggering and time base

Summary of Contents for HM303-6

Page 1: ...O s c i l l o s c o p e H M 3 0 3 6 Manual English...

Page 2: ...lifiers 12 X Y Operation 13 Phase comparison with Lissajous figures 13 Phase difference measurement 13 in DUAL mode 13 Phase difference measurement in DUAL mode 14 Measurement of an amplitude modulati...

Page 3: ...effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT gap As the bandwidth of each amplifier stage is higher than the tot...

Page 4: ...oise Amplifiers Time Base 0 2 s 100 ns cm with X Magnification to 10 ns cm Triggering from 0 to 50 MHz from 5 mm signal level 100 MHz 8 mm Analog mode provides unexcelled signal presentation at high r...

Page 5: ...1 s div 1 2 5 Sequence Accuracy 3 Variabel uncalibrated 2 5 1 to 0 5 s div X Magnification x 10 up to 10 ns div Accuracy 5 Hold Off Time variable to approx 10 1 XY Bandwidth X Amplifier 0 2 5 MHz 3 d...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...e atmosphere or if there is danger of explosion The operating position may be any however sufficient ventilation mustbeensured convectioncooling Prolongedoperationrequires the horizontal or inclined p...

Page 8: ...83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective value Vp simple p...

Page 9: ...frequency Depending on the time base setting of the TIME DIV switch one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div ms div and s div...

Page 10: ...d The attenuator switch must then be turned back to the left until the vertical signal height is only 3 8div With a signal amplitude greater than 160Vpp an attenuator probe must be inserted before the...

Page 11: ...be visible after a short warm up period of approx 10 seconds Adjust Y POS I and X POS controls to center the baseline Adjust INTENS intensity and FOCUS controls for medium brightness and optimum sharp...

Page 12: ...is slightly more difficult but causes a better result The rising edge should be as steep as possible with a pulse top remaining as straight and horizontal as possible After completion of the HF adjus...

Page 13: ...uite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no influence on the result The following must be noted...

Page 14: ...tion F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the two values a and b are read from the screen the modulation factor is calculated from where a UT 1 m and b UT 1 m The variable controls for amplitude and...

Page 15: ...a positive potential That has nothing to do with zero or ground potential and absolute voltage values The positive slope may also lie in a negative part of a signal A falling edge sign triggers when...

Page 16: ...ossible in this trigger mode External triggering The internal triggering is disconnected by depressing the TRIG EXT button The time base can be triggered externally via the TRIG EXT socket using a 0 3...

Page 17: ...are required The test leads are connected to the insulated socket and the adjacent ground socket beneath the screen The component can be connected to the test leads either way round After use to retu...

Page 18: ...can be identified by comparison with a known diode Testing Transistors Three different tests can be made to transistors base emitter base collector and emitter collector The resulting test patterns ar...

Page 19: ...soldered This terminal should then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which i...

Page 20: ...II display SLOPE Selects the slope of the trigger signal Element Function SLOPE rising edge pushbutton switch falling edge TR LED lights if sweep is triggered LED LEVEL Adjustment of trigger level kno...

Page 21: ...ast 2 5 fold For time measurements turn to right hand stop Element Function XY Selects X Y operation pushbutton switch stops sweep X signal via CH I Attention Phosphor burn in without signal TRIG EXT...

Page 22: ...ggering TRIG EXT pushbutton depressed sync signal 0 3Vpp to 3Vpp to TRIG EXT socket Line triggering normal triggering AT NM and ALT pushbuttons depressed Select trigger coupling with TRIG MODE selecto...

Page 23: ...M A N U A L H A N D B U C H M A N U E L...

Page 24: ...stra e 6 A Rohde Schwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes S...

Reviews: