Hameg HM303-6 Manual Download Page 10

10

Subject to change without notice

When investigating pulse or square waveforms, the critical
feature is the 

risetime of the voltage step

. To ensure that

transients, ramp-offs, and bandwidth limits do not unduly influence
the measuring accuracy, the risetime is generally measured
between 

10%

 and 

90%

 of the vertical pulse height. For

measurement adjust the Y attenuator switch with its variable
control together with the 

Y-POS.

 control so that the pulse height

is precisely aligned with the 0% and 100% lines of the internal
graticule. The 10% and 90% points of the signal will now coincide
with the 10% and 90% graticule lines. 

The risetime is given by

the product of the horizontal distance in div. between these
two coincidence points and the time coefficient setting

. If X

x10 magnification is used, this product must be divided by 10. The

fall time

 of a pulse can also be measured by using this method.

The following figure shows correct positioning of the oscilloscope
trace for accurate risetime measurement.

With a time coefficient of 0.2µs/div. and pushed X-MAG x10
button the example shown in the above figure results in a
measured total risetime of

t

tot

 = 1.6div x 0.2µs/div. / 10 = 

32ns

When very fast risetimes are being measured, the risetimes of
the oscilloscope amplifier and of the attenuator probe has to be
deducted from the measured time value. The risetime of the

signal can be calculated using the following formula.

In this t

tot

 is the total measured risetime, t

osc

 is the risetime of the

oscilloscope amplifier (approx. 10ns), and tp the risetime of the
probe (e.g. = 2ns). If t

tot

 is greater than 100ns, then t

tot

 can be

taken as the risetime of the pulse, and calculation is unnecessary.

Calculation of the example in the figure above results in a signal
risetime

t

r

32

2

  - 10

2

 - 2

 = 

30,3ns

The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to
measure in any display position and at any signal amplitude. It
is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness and
that the horizontal distance at 10% and 90% of the amplitude
is measured. If the edge shows rounding or overshooting, the
100% should not be related to the peak values but to the mean
pulse heights. Breaks or peaks (glitches) next to the edge are
also not taken into account. With very severe transient
distortions, the rise and fall time measurement has little meaning.
For amplifiers with approximately constant group delay
(therefore good pulse transmission performance) the following
numerical relationship between rise time 

tr 

(

in ns

) and bandwidth

B

 (

in MHz

) applies:

Connection of Test Signal

Caution:
When connecting unknown signals to the oscilloscope
input, always use automatic triggering and set the DC-AC
input coupling switch to AC. The attenuator switch should
initially be set to 20V/div.

Sometimes the trace will disappear after an input signal has
been applied. The attenuator switch must then be turned back
to the left, until the vertical signal height is only 3-8div. With
a signal amplitude greater than 160V

pp

, an attenuator probe

must be inserted before the vertical input. If, after applying
the signal, the trace is nearly blanked, the period of the signal
is probably substantially longer than the set value on the

TIME/DIV.

 switch. It should be turned to the left to an

adequately larger time coefficient.

The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as HZ
32 or HZ 34, or reduced through a x10 or x100 attenuator
probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to approx.
50kHz). For higher frequencies, the signal source must be of
low impedance, i.e. matched to the characteristic resistance
of the cable (as a rule 50

). Especially when transmitting

square and pulse signals, a resistor equal to the characteristic
impedance of the cable must also be connected across the
cable directly at the Y-input of the oscilloscope. When using
a 50

 cable such as the HZ 34, a 50

 through termination type

HZ22 is available from 

HAMEG

.

When transmitting square signals with short rise times, transient
phenomena on the edges and top of the signal may become
visible if the correct termination is not used. A terminating
resistance is sometimes recommended with sine signals as
well. Certain amplifiers, generators or their attenuators maintain
the nominal output voltage independent of frequency only if
their connection cable is terminated with the prescribed
resistance. Here it must be noted that the terminating resistor
HZ22 will only dissipate a maximum of 2 Watts. This power is
reached with 10 Vrms or  at 28.3 V

pp

 with sine signal.

If a x10 or x100 attenuator probe is used, no termination is
necessary. In this case, the connecting cable is matched
directly to the high impedance input of the oscilloscope. When
using attenuators probes, even high internal impedance sources
are only slightly loaded (approx. 10M

 II 16pF or 100M

 II 9pF

with HZ53). Therefore, if the voltage loss due to the attenuation
of the probe can be compensated by a higher amplitude setting,
the probe should always be used. The series impedance of the
probe provides a certain amount of protection for the input of
the vertical amplifier. Because of their separate manufacture,
all attenuator probes are only partially compensated, therefore
accurate compensation must be performed on the oscilloscope
(see ”Probe compensation” ).

Standard attenuator probes on the oscilloscope normally
reduce its bandwidth and increase the rise time. In all cases
where the oscilloscope bandwidth must be fully utilized (e.g.
for pulses with steep edges) we strongly advise using the

probes HZ51

 (x10) 

HZ52

 (x10 HF) and 

HZ54

  (x1 and x10). This

can save the purchase of an oscilloscope with larger bandwidth
and has the advantage that defective components can be
ordered from 

HAMEG

 and replaced by oneself. The probes

mentioned have a HF-calibration in addition to low frequency
calibration adjustment. Thus a group delay correction to the

Type of signal voltage

Summary of Contents for HM303-6

Page 1: ...O s c i l l o s c o p e H M 3 0 3 6 Manual English...

Page 2: ...lifiers 12 X Y Operation 13 Phase comparison with Lissajous figures 13 Phase difference measurement 13 in DUAL mode 13 Phase difference measurement in DUAL mode 14 Measurement of an amplitude modulati...

Page 3: ...effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT gap As the bandwidth of each amplifier stage is higher than the tot...

Page 4: ...oise Amplifiers Time Base 0 2 s 100 ns cm with X Magnification to 10 ns cm Triggering from 0 to 50 MHz from 5 mm signal level 100 MHz 8 mm Analog mode provides unexcelled signal presentation at high r...

Page 5: ...1 s div 1 2 5 Sequence Accuracy 3 Variabel uncalibrated 2 5 1 to 0 5 s div X Magnification x 10 up to 10 ns div Accuracy 5 Hold Off Time variable to approx 10 1 XY Bandwidth X Amplifier 0 2 5 MHz 3 d...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...e atmosphere or if there is danger of explosion The operating position may be any however sufficient ventilation mustbeensured convectioncooling Prolongedoperationrequires the horizontal or inclined p...

Page 8: ...83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective value Vp simple p...

Page 9: ...frequency Depending on the time base setting of the TIME DIV switch one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div ms div and s div...

Page 10: ...d The attenuator switch must then be turned back to the left until the vertical signal height is only 3 8div With a signal amplitude greater than 160Vpp an attenuator probe must be inserted before the...

Page 11: ...be visible after a short warm up period of approx 10 seconds Adjust Y POS I and X POS controls to center the baseline Adjust INTENS intensity and FOCUS controls for medium brightness and optimum sharp...

Page 12: ...is slightly more difficult but causes a better result The rising edge should be as steep as possible with a pulse top remaining as straight and horizontal as possible After completion of the HF adjus...

Page 13: ...uite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no influence on the result The following must be noted...

Page 14: ...tion F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the two values a and b are read from the screen the modulation factor is calculated from where a UT 1 m and b UT 1 m The variable controls for amplitude and...

Page 15: ...a positive potential That has nothing to do with zero or ground potential and absolute voltage values The positive slope may also lie in a negative part of a signal A falling edge sign triggers when...

Page 16: ...ossible in this trigger mode External triggering The internal triggering is disconnected by depressing the TRIG EXT button The time base can be triggered externally via the TRIG EXT socket using a 0 3...

Page 17: ...are required The test leads are connected to the insulated socket and the adjacent ground socket beneath the screen The component can be connected to the test leads either way round After use to retu...

Page 18: ...can be identified by comparison with a known diode Testing Transistors Three different tests can be made to transistors base emitter base collector and emitter collector The resulting test patterns ar...

Page 19: ...soldered This terminal should then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which i...

Page 20: ...II display SLOPE Selects the slope of the trigger signal Element Function SLOPE rising edge pushbutton switch falling edge TR LED lights if sweep is triggered LED LEVEL Adjustment of trigger level kno...

Page 21: ...ast 2 5 fold For time measurements turn to right hand stop Element Function XY Selects X Y operation pushbutton switch stops sweep X signal via CH I Attention Phosphor burn in without signal TRIG EXT...

Page 22: ...ggering TRIG EXT pushbutton depressed sync signal 0 3Vpp to 3Vpp to TRIG EXT socket Line triggering normal triggering AT NM and ALT pushbuttons depressed Select trigger coupling with TRIG MODE selecto...

Page 23: ...M A N U A L H A N D B U C H M A N U E L...

Page 24: ...stra e 6 A Rohde Schwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes S...

Reviews: