Chapter 7. Troubleshooting
EQ-99X Operation & Maintenance Manual
41
of
45
7.1 Issues and Remedies
See below for a list of possible issues that may be encountered while operating the EQ-99X system
and the potential
steps to resolve those issues.
LED Light Name
and Status
Possible Issues
Troubleshooting Steps
LAMP MODULE
FAULT is
ON
One of the following has occurred in the lamp head:
CONTROLLER
FAULT is
ON
One of the following has occurred in the power
supply controller:
Possible Issues and Remedies
Table 10.
1. A control cable is not connected.
2.
The lamp head has overheated (above 85 °C).
3. The laser subminiature assembly (SMA) is not
connected.
4. The lamp bulb is not properly installed.
5. There was a failure to ignite the plasma
inside the lamp bulb, or the plasma inside the
lamp bulb has been extinguished after
ignition. This also causes the LAMP ON light
to turn
OFF
and LASER ON light to turn
OFF
.
1. Clear the fault lights by toggling
the EQ-99-RC Remote Control
between the ON and OFF
positions. The EQ-99X system
will
not start if a fault condition exists.
2. Confirm that the black jumper
plug is fully inserted into the back
of the EQ-99-RC Remote Control
(or, if using another remote
control, confirm that the external
interlock contact is closed).
3. Check that the bulb and laser fiber
optic cable are properly connected
to the lamp head.
4. If either the LAMP MODULE FAULT
or CONTROLLER FAULT light will
still not clear, contact Energetiq
.
1. An external interlock is open.
2. The power supply controller printed circuit
board has overheated (above 82 °C).
3. The laser has overheated (above 66 °C).
4. There is a regulation error with the laser
current.
5. There is a regulation error with the laser
temperature.
6.
There is a laser failure.