![Hamamatsu ENERGETIQ LDLS EQ-99X Operation And Maintenance Manual Download Page 13](http://html1.mh-extra.com/html/hamamatsu/energetiq-ldls-eq-99x/energetiq-ldls-eq-99x_operation-and-maintenance-manual_3392253013.webp)
Chapter 2. System Description
EQ-99X Operation & Maintenance Manual
13
of
45
LED Label
Meaning (when lit)
LAMP MODULE FAULT
One of the following has occurred in the lamp head:
B. Input/Output (I/O) Port
The I/O port is used to connect a remote control to the EQ-99X system. You can either use the EQ-99-RC Remote
Control provided with the system, or another remote control. See
4.4.2 Installing Alternative Remote Control
for pin
assignments and functions.
C. Power Port
This is a latching port for 12VDC power. Power can be provided using the included 12VDC power supply. Alternatively,
the EQ-99X system
can be powered directly from a customer-provided 12VDC power supply. See
D. Lamp Head Signal Port
This port is used to provide various power and control signals to/from the lamp head. No other connector or cable
may be used with the EQ-99X system
other than the one supplied.
E. Laser Fiber Optic Cable
The laser light is delivered from the power supply controller to the lamp head via a fiber optic cable with armored
protection. The fiber is permanently attached to the power supply controller and connects to the lamp head via an
SMA-type connector.
It is critical that this armored fiber be treated with care and inspected for any abnormalities prior to operation. Avoid
sharp bends which can permanently damage the fiber. The minimum bend radius is 30 mm (1.18 in). Avoid crimping
for more information on installation and
for more information on fiber end-face cleaning.
1. Control cable not connected properly
2. Lamp head internal temperature too high
3. Laser fiber not correctly connected
4. Lamp bulb not correctly installed
5. Ignition failure