13. TROUBLSHOOTING
If an abnormality occurs, look up the possible causes in the following tables and, if necessary, report
the details to Hamamatsu subsidiary or your local distributor.
•
When the following measures are taken, turn off the camera and peripherals
and secure a space where you can work stably and do it without risks.
•
Do not attempt to dismantle or modify the camera. Do not touch parts that are
not indicated in this manual.
13-1 TURNING ON THE POWER
Symptom
Cause
Measures
Chapter
Image is not transferred
AC adapter is unconnected to an
electrical outlet
Connect the AC adapter
surely
AC adapter or other cable is loose
Reconnect the cable and
tighten the fixing screws of
the connector surely
AC adapter or other cable is broken
Exchange it for a new cable
The fan inside the camera does
not rotate (at Air-cooling)
The cooling method of DCAM
Configurator is set to “Water-cooling”
Change the setting to
“Air-cooling”
13-2 ALTHOUGH IMAGES ARE TRANSFERRED
Symptom
Cause
Measures
Chapter
Scratches or discoloration
visible on the screen
Lens is dirty
Wipe the lens.
Image is blurred
Lens is not focused
Contact a Hamamatsu
subsidiary or your local
distributor.
Only shadowed images are
output
Lens mount cap has been left on
Remove the cap.
Amount of light is too much or too low
Adjust amount of light.
All screens overflow
Too much amount of light
Reduce amount of light.
Noise appears on the screen
Ambient temperature is high
Lower the temperature to
25 °C.
Change the cooling method to
Water-cooling.
Exogenous noise
Find and remove cause.
Poor connection of internal connector
Contact a Hamamatsu
subsidiary or your local
distributor.
Defective circuit system
Ring a buzzer (beep tone)
The temperature inside the camera
rises abnormality
As soon as turn off the power
switch, unplug the AC supply.
Wait till the temperature falls,
and then remove the cause of
the overheating. After that turn
on the power supply of the
camera again. If the symptom
is unimproved, and the cause
is unidentifiable, contact a
Hamamatsu.