14. SPECIFICATIONS
(1) Performances
Imaging device
Scientific CMOS image sensor
Effective number of pixels
4608 (H) × 2592 (V)
Cell size
5.5 μ
m (H) × 5.5 µm (V)
Effective area
25.344 mm (H) × 14.256 mm (V)
Cooling method
Peltier Forced air-cooling, Water-cooling
Cooling temperature
at Forced air-cooling +20 °C (Ambient temperature: +25 ºC)
at Water-cooling
+20 °C (Water temperature: +25 ºC)
Dark current *
1
15 electrons/pixel/s
Dark Signal Non-Uniformity (DSNU)
*
1
0.5 electrons r.m.s.
Photo Response Non-Uniformity
(PRNU) *
1
3 % or less (20 000 electrons signal)
Linearity error *
1
EMVA 1288 standard
1 % or less
< 500 electrons signal
1 % or less
Dark offset
200 counts (at Normal (1×1) readout)
Exposure time
6.304
μ
s to 1 s (in step of 6.304
μ
s)
Standard Full Well Capacity
mode
High Full Well Capacity
mode
Full well capacity *
1
1000 electrons
38 000 electrons
Dynamic range *
2
650:1
17 000:1
Readout noise *
1
2.0 electrons (r.m.s.)
1.5 electrons (median)
2.7 electrons (r.m.s.)
2.2 electrons (median)
Conversion factor *
1
0.26 electrons/count
0.58 electrons/count
Digital output
12 bit
16 bit
Maximum pixel rate
1445 Mpixels/s
358 Mpixels/s
Frame rate
4608 × 2592
121 fps
30 fps
2048 × 2048
307 fps
38 fps
1024 × 1024
610 fps
76 fps
512 × 512
1201 fps
150 fps
128 × 128
4406 fps
550 fps
*
1
Typical value
*
2
Calculated from the ratio of the full well capacity and the readout noise.
•
The frame rates in the table above are the output frame rates from the camera.
•
Since the frame rates of image acquisition by software depend on the PC performance and
the application software, they may be slower than the frame rates in the table above without
sufficient PC performance and application software. The slower frame rates mean the lost
frames.
•
The lost frames will be reduced by using Frame bundle function of DCAM-API software that
transfers multiple frames at once.