C13440-20CU / C13440-20CU01 Instruction manual_Ver.1.3
71
11-4-4 READOUT TIME OF THE HORIZONTAL LINE
Readout time and exposure time can be varied with Dual Lightsheet Readout Mode for synchronizing
the camera readout with the illumination scan.
Vn = Number of vertical line
1H = 9.744 36 µs to 100 ms
Readout time = Vn × 1H
The range of exposure time with Lightsheet Readout Mode is as shown below.
The maximum exposure time can be decided according to the readout time of a frame, which is longer
than 10 s or not.
1H to (Vn × 1H) : when (Vn × 1H) < 10 s
1H to 10 s
: when (Vn × 1H) ≥ 10 s
11-4-5 TIMING DIAGRAM
(1) Free running mode
Figure 11-35 (Ex. Top: top to center / Bottom: center to bottom)
0H
1H
・
・
1023H
0H
・
・
1022H
1023H
Internal exposure timing setting
Exposure
Sensor readout
USB 3.0 data output
Camera Link data output
Exposure
1H can be
varied