h. Power switch [POWER]
The power is turned on/off.
- When the power switch is set to "ON", the camera turns on and starts initialization and the lamp
blinks in green.
- When the initialization is completed, the lamp color stays in green.
- When the camera transfers data, and the lamp color turns orange.
- When the power switch is set to "OFF", the camera returns to the power off state and the lamp turns
off.
i. DC power input connector [DC IN]
This is the power supply terminal. Use the accessory AC adapter.
j. USB 3.0 interface connector [USB 3.0]
This is connected to the USB 3.0 interface connector on the computer.
•
When a connection interface is changed from Camera Link to USB 3.0, and vice
versa, the application software must be closed and the camera must be turned off.
•
Do not connect Camera Link and USB 3.0 interface simultaneously.
k. Air inlet
This is the inlet for the heat ventilation.
•
If the camera is being operated in an enclosed environment, ensure to keep
clearance at least 10 cm from both intake and exhaust vents when setting up.
•
To prevent overheating inside the camera, do not wrap the camera in cloth or other
material, or block the camera’s ventilation.
l. Base plate
This base plate is for attaching to a lab jack or a tripod.
When you use an adjustable pole (A11185-01) or a fixing bracket (A13261-01), please detach the base
plate, and then attach the options.
Note
•
Please refer to the installation manual of options about the detaching method of
base plate and the attaching method of the option.