C13440-20CU / C13440-20CU01 Instruction manual_Ver.1.3
16
•
Place the camera the water connectors to be lateral side. Do not place the rear
panel of the camera, which connectors are located, to be at the bottom (Do not
block ventilation openings.).
a. Lens mount
C13440-20CU can be attached to C-mount lens or an optics system.
C13440-20CU01 can be attached to F-mount lens or an optics system.
Note
•
The depth of the C-mount is 7 mm. Screwing in the C-mount too deeply might
scratch the glass surface.
b. WATER connector [WATER] (when using Water-cooling)
It connects the camera and the circulating water cooler with the cooling water hoses. The connector
position of WATER IN/OUT is not specified.
•
See 8 “WATER COOLING” for instruction of water-cooling.
c. Camera Link interface connector 1 [DIGITAL OUT 1]
d. Camera Link interface connector 2 [DIGITAL OUT 2]
The connector 1 is connected to the Camera Link interface connector 1 on the computer.
The connector 2 is connected to the Camera Link interface connector 2 on the computer.
•
When a connection interface is changed from Camera Link to USB 3.0, and vice
versa, the application software must be closed and the camera must be turned off.
•
Do not connect Camera Link and USB 3.0 interface simultaneously.
e. Trigger input connector [EXT.TRIG]
This is used when the camera is being operated using external synchronization.
Input is 3.3 V LVCMOS level, and input impedance
is 10 kΩ.
When an external trigger is input, the trigger is activated at the falling or rising edge of the signal. (You
can choose external trigger polarity between Negative and Positive.)
f. Timing out connector 1,2,3 [TIMING 1,2,3]
This is used when peripheral device(s) require synchronization with the camera.
Output is 3.3 V LVCMOS level, and it is output though
BUS TRANSCEIVER IC SN74LVC541
. Output
impedance is 33 Ω.
Note
•
Determine termination according to cable length and so on.
g. STATUS lamp [STATUS]
The LED indicates status of camera.
Lighting color
Status of power distribution
Turn off (no color)
Power off
Orange (Blinking)
Initialization
Green
(lighting)
Power on
Orange (lighting)
Data transfer
Red
(lighting)
Heat up
•
When the camera heats up, stop operation and unplug the
AC adapter immediately.