Troubleshooting

The following table of potential indicators provides a quick reference of the descriptions, probable causes 
and most common solutions regarding the operation of the HaloFogger® EXT. If the information in this 
table does not describe or solve the problem you are having, contact Customer Service at 1 302-454-8102, 
Monday through Friday, 9:00AM to 5:00PM EST. 

Symptom

Possible Cause

Solution

No mist is being 
produced

The nozzle is clogged 
misaligned.

Remove the nozzle and clean according to 
the instructions in this manual.

The disinfectant fluid
tank is empty.

Check the fluid level indicator lights. If the 
orange light is illuminated, follow the fill 
instructions in this manual to refill the tank.

Start button not pressed.

Green indicator light is 
blinking

Reservoir Door is not  
closed correctly.

Close Reservoir Door making sure it “clicks”  
when closed.

Magnet in Reservoir Door 
is not in place.  

Call Customer Service.

No power 

The power cord is not 
plugged in or the power 
source is not on.

Check to make sure that the power cord is 
plugged into both the fogger and the wall 
outlet. Then, check to see that the outlet is 
working. The green status indicator light 
should be illuminated when properly set up. 
 

The fuse is blown.

Check to see if the fuse is blown. Fuse (and 
a spare) is located behind a discreet panel 
just below where the power cord connects 
to the HaloFogger EXT. 

Knocking noise

The compressor is not 
running properly.

Call Customer Service.

Disinfecting procedure 
is interrupted

Power loss during 
treatment process 
or other unforeseen 
circumstances

.

Confirm power has been restored. Re-
enter room after hydrogen peroxide 
concentration level is determined to be at 
or below 1 ppm. Restart treatment process 
from beginning. 

HaloFogger® EXT Operation Manual_1911

Extended hose assembly 
is not connected correctly.

Check both couplings to verify proper 
connection. An audible ‘click’ should be 
heard when the couplings are connected 
correctly.

Press and hold the Start button for 2 
seconds to begin fogging.

The unit not plugged in.

Check to make sure that the power cord 
is plugged into both the HaloFogger and 
a wall outlet. Then, check to see that the 
wall outlet is working. The green status 
indicator light should be illuminated when 
properly set up.

Summary of Contents for HaloFogger EXT

Page 1: ...READ AND SAVE THESE INSTRUCTIONS Operation Manual 110V 220V EXT 1911 EXT...

Page 2: ...n and dispensing disinfectant The room should be unoccupied Large Wheels For easy maneuvering Power Cord Connection Reservoir Drain High Precision Nozzle Dry mist of disinfectant is dispensed through...

Page 3: ...lashing or overfilling which could harm the user and or the device 12 Halosil International is not responsible for any injury or damage caused by using this device outside of the specific parameters d...

Page 4: ...ger EXT without the extended nozzle hose assembly fully connected Damage will occur Filling Instructions 1 Open the HaloFogger EXT Reservoir Door by lifting front edge of door 2 Locate the disinfectan...

Page 5: ...r wheel 4 Take the clear tubing provided and drop one end into the waste bottle 5 Working the adjustable ball joints of the nozzle assembly mounting arm point the nozzle straight down Then move the no...

Page 6: ...determine if the measured H2O2 level has dropped to about 1 ppm or below C If no written protocol has been established and you have no H2O2 sensor assume a waiting time of AT LEAST 110 MINUTES AFTER F...

Page 7: ...r data logger via Bluetooth The HaloLog App does not need to be connected to a fogger before a treatment run is started The on board data logger captures all of the operating data which can be reviewe...

Page 8: ...ll be displayed 4 To initiate a connection tap on any HaloFogger serial number displayed on the Halo List Screen When the connection is made the Run Events Screen will appear The top row of the screen...

Page 9: ...and total Run Time for that HaloFogger highlighted in GREEN The HaloLog data stored on the selected HaloFogger will download automatically after the Bluetooth handshake has been established which is...

Page 10: ...nts will identified BILLABLE as a default Events that are not treatments a test run for example can be identified as NON BILLABLE These events will not appear in the downloadable CSV file but will be...

Page 11: ...ddress A keyboard will appear to enter address ACTION Tap to email data The iDevice s email screen will open to confirm message ACTION Tap SEND in upper right corner of screen to send message Back up...

Page 12: ...o a HaloFogger 2 In the event that HaloLogs of the same HaloFogger have been downloaded to multiple devices If one user has made changes to the HaloLog for example added room building or operator info...

Page 13: ...remove Step 3 Separate Parts Remove Main Bolt from Nozzle Insert Soak both parts in soap and water Rinse parts thoroughly with water Step 4 Reassemble Nozzle Carefully insert the Main Bolt back into...

Page 14: ...power cord is plugged into both the fogger and the wall outlet Then check to see that the outlet is working The green status indicator light should be illuminated when properly set up The fuse is blow...

Page 15: ...sembly 0750 000 includes retainer cap nozzle insert main bolt and O ring Re Order Number Part Part Number 110V HaloFogger EXT US and Canada 0753 110EXT 220V HaloFogger EXT 0753 220EXT HaloMist All Pur...

Page 16: ...of room being treated CAUTION PRECAUCION DO NOT ENTER ENTERINO Room Fogging in Progress Treatment Date Start Time Estimated Room Re entry Time Refer to HaloMist Use Instructions to determine room re...

Page 17: ...was purchased and the unit s serial number Proof of purchase may be required If Halosil International determines that the fogger qualifies for warranty coverage a Return Merchandise Authorization RMA...

Page 18: ...uce infection rates at your facility by visiting www halosil com There you can find out more about the Halo Disinfection System download literature and case studies and find a distributor near you 91...

Reviews: