Getting Started with the HaloFogger® EXT

I

MPORTANT

Before using your HaloFogger, it is important that you:
1. Plug the HaloFogger EXT into a power source for 24 hours to fully charge the HaloLog’™ 
Data Logger’s internal back-up battery.
2. Download the HaloMist™ All-Purpose Disinfectant Use Instructions. 
Go to 

www.halosil.com

, register your new fogger purchase to download.

Setup Instructions
Locate the connector end of the power cord. Insert the connector into the power receptacle located at the 
bottom of the rear panel of the HaloFoggerEXT. Ensure that the connector is securely in place. Plug the 
HaloFogger EXT into a standard electrical outlet. The green status indicator light will illuminate.

Setup and Filling
THE HALOFOGGER EXT IS DESIGNED FOR USE WITH HALOMIST™ ALL-PURPOSE DISINFECTANT AND SHOULD 
NOT BE OPERATED WITH ANY OTHER MANUFACTURERS DISINFECTING OR CLEANING PRODUCT.

Installing Nozzle Assembly on the HaloFogger EXT
Locate the Nozzle Assembly from the HaloFogger EXT shipping box and unwrap. The assembly includes 
the high precision nozzle head, adjustable mounting arm, and extra long hose with quick connect hose 
couplings. Locate the graphic icon on each quick connect coupling, match with corresponding icon and 
bulkhead fitting on the top panel of the HaloFogger, and press firmly to engage connector until an audible 
‘click’ is heard.. 

(Press button on coupling to disconnect.)

WARNING
Do not operate the HaloFogger EXT without the extended nozzle hose assembly fully connected. 
Damage will occur.

Filling Instructions
1. Open the HaloFogger EXT Reservoir Door by lifting front edge of door.
2. Locate the disinfectant fluid level indicator lights to the left of the funnel. One flashing red light will be 
illuminated indicating the HaloFogger EXT is empty.
3. Use personal protection equipment (PPE) including wrap-around style goggles and gloves when pouring 
disinfectant fluid into the device.
4. Open a gallon (3.8 liters) bottle of HaloMist™ Disinfectant and SLOWLY pour the contents into the funnel 
area. AVOID SPLASHING.
5. The fluid level indicator lights will illuminate as you pour to show how much disinfectant is in the device.
6. The HaloFogger EXT holds two gallons (7.6 liters) of disinfectant when full. Stop pouring when the top 
green light on the fluid level indicator is illuminated. Do not overfill.
7. Wipe up any spills with a clean paper towel or rag. Thoroughly rinse off gloves and hands.
8. Close Reservoir Door.
9. The HaloFogger EXT is now ready to be used.

Basic Operating Instructions

1. Set the timer to the number of minutes specified for your room size. See Quick Instructions on 
rear of unit.
2. Plug the HaloFogger EXT into a standard, grounded electrical wall outlet. When the power connection has 
been made, the green status indicator light on top of the device will illuminate.
3. Check fluid level indicator lights. It is recommended the unit’s reservoir be topped off before starting the 
treatment process. Refer to “Getting Started with the HaloFogger EXT” above to add/refill disinfectant.

HaloFogger® EXT Operation Manual_1911

Summary of Contents for HaloFogger EXT

Page 1: ...READ AND SAVE THESE INSTRUCTIONS Operation Manual 110V 220V EXT 1911 EXT...

Page 2: ...n and dispensing disinfectant The room should be unoccupied Large Wheels For easy maneuvering Power Cord Connection Reservoir Drain High Precision Nozzle Dry mist of disinfectant is dispensed through...

Page 3: ...lashing or overfilling which could harm the user and or the device 12 Halosil International is not responsible for any injury or damage caused by using this device outside of the specific parameters d...

Page 4: ...ger EXT without the extended nozzle hose assembly fully connected Damage will occur Filling Instructions 1 Open the HaloFogger EXT Reservoir Door by lifting front edge of door 2 Locate the disinfectan...

Page 5: ...r wheel 4 Take the clear tubing provided and drop one end into the waste bottle 5 Working the adjustable ball joints of the nozzle assembly mounting arm point the nozzle straight down Then move the no...

Page 6: ...determine if the measured H2O2 level has dropped to about 1 ppm or below C If no written protocol has been established and you have no H2O2 sensor assume a waiting time of AT LEAST 110 MINUTES AFTER F...

Page 7: ...r data logger via Bluetooth The HaloLog App does not need to be connected to a fogger before a treatment run is started The on board data logger captures all of the operating data which can be reviewe...

Page 8: ...ll be displayed 4 To initiate a connection tap on any HaloFogger serial number displayed on the Halo List Screen When the connection is made the Run Events Screen will appear The top row of the screen...

Page 9: ...and total Run Time for that HaloFogger highlighted in GREEN The HaloLog data stored on the selected HaloFogger will download automatically after the Bluetooth handshake has been established which is...

Page 10: ...nts will identified BILLABLE as a default Events that are not treatments a test run for example can be identified as NON BILLABLE These events will not appear in the downloadable CSV file but will be...

Page 11: ...ddress A keyboard will appear to enter address ACTION Tap to email data The iDevice s email screen will open to confirm message ACTION Tap SEND in upper right corner of screen to send message Back up...

Page 12: ...o a HaloFogger 2 In the event that HaloLogs of the same HaloFogger have been downloaded to multiple devices If one user has made changes to the HaloLog for example added room building or operator info...

Page 13: ...remove Step 3 Separate Parts Remove Main Bolt from Nozzle Insert Soak both parts in soap and water Rinse parts thoroughly with water Step 4 Reassemble Nozzle Carefully insert the Main Bolt back into...

Page 14: ...power cord is plugged into both the fogger and the wall outlet Then check to see that the outlet is working The green status indicator light should be illuminated when properly set up The fuse is blow...

Page 15: ...sembly 0750 000 includes retainer cap nozzle insert main bolt and O ring Re Order Number Part Part Number 110V HaloFogger EXT US and Canada 0753 110EXT 220V HaloFogger EXT 0753 220EXT HaloMist All Pur...

Page 16: ...of room being treated CAUTION PRECAUCION DO NOT ENTER ENTERINO Room Fogging in Progress Treatment Date Start Time Estimated Room Re entry Time Refer to HaloMist Use Instructions to determine room re...

Page 17: ...was purchased and the unit s serial number Proof of purchase may be required If Halosil International determines that the fogger qualifies for warranty coverage a Return Merchandise Authorization RMA...

Page 18: ...uce infection rates at your facility by visiting www halosil com There you can find out more about the Halo Disinfection System download literature and case studies and find a distributor near you 91...

Reviews: