MP5A Operation and Maintenance Manual
________________________________________________________________________________________________________________
Copyright
©
2023
‐
2024
Halogen
Systems,
Inc.
All
rights
reserved
worldwide.
Rev.:
4.02
26
Legend
for
display
locations
Page
2
pH
Signal
pH
Reference
Signal
Sensor
Serial
Number
Number
of
Hours
of
Operation
Sensor
Serial
Number
is
helpful
when
requesting
technical
support.
Hours
of
operation
is
an
indication
of
when
wear
parts
need
to
be
replaced.
8.7.
Problems not resulting in an Error Code
8.7.1
Measured water or sensor reading is fluctuating and it is difficult to calibrate sensor
Put
sensor
in
a
bucket
with
sampled
water.
Take
a
grab
sample
and
calibrate
the
sensor.
Calibrate
the
sensor
and
replace
into
process.
8.7.2
Chlorine and or pH reading is erratic
This
is
often
caused
by
air.
To
check
the
sensor
operation,
place
it
is
a
bucket
with
sampled
water
and
determine
in
the
value
measured
is
stable
over
several
measurements.
If
so,
then
the
problem
is
likely
the
installation.
Air
may
be
present
in
the
sampling
stream,
or
the
sensor
orientation
may
be
wrong.
Sensor
is
covered
by
brown
slime
or
other
contaminants
Clean
sensor
using
one
of
the
two
methods
described
in
the
Maintenance
Section
in
the
Appendix.
8.7.3
PLC input does not match the output of the Display
See
the
Section
on
Calibration
of
the
4
‐
20
mA
Output
in
the
Appendix.
8.7.4
Immersed sensor is installed in a basin which is drained periodically
Sensor
Operation
can
be
stopped
without
powering
off
by
using
the
STOP
Function
below:
8.7.5
Stop Function
A
STOP
command
will
halt
measurement
until
the
system
is
restored.
Pressing
MENU
and
BACK
at
the
same
time
will
enable
the
STOP
cycle.
Repeating
the
sequence
will
restart
the
cycle.
9.
Section Replacement Parts and Accessories
Wear
Part
Replacement
Quantity
PN
Description
Frequency
1
WP
‐
NH
Wear
Part
Replacement
Kit
including:
12
Months
1
NCAP
‐
0001
Sensor
End
Cover
Assembly
1
IMPA
‐
0001
Impeller
2
CLBL
‐
0001
Bag
of
15x
1/8"
Acrylic
cleaning
beads
1
MTRR
‐
0001
Motor
Wear
Ring
2
FSTR
‐
0005
Sensor
Cover
Screws:
No.
4
5/8"
316
SS
Self
Tapping
Phillips
1
KITS
‐
0001
pH
Sensor
Replacement
Kit
including:
12
Months
1
PHSA
‐
0001
pH
Reference
Electrode
Assembly
1
TOOL
‐
0001
Tool
‐
pH
sensor
removal
1
ORNG
‐
0004
O
‐
ring
‐
#12
NSF61,
EPDM,
ID:
9.25
mm,
CS:
1.78
mm