Halogen MP5A Operation & Maintenance Manual Download Page 20

MP5A Operation and Maintenance Manual

 

________________________________________________________________________________________________________________

 

Copyright

 

©

 

2023

2024

 

Halogen

 

Systems,

 

Inc.

 

All

 

rights

 

reserved

 

worldwide.

  

Rev.:

 

4.02

 

 

20

 

7.

 

Maintenance  

7.1.

 

Wastewater Maintenance 

The

 

Halogen

 

MP5

 

sensor

 

is

 

very

 

low

 

maintenance

 

and

 

can

 

be

 

installed

 

in

 

either

 

Wastewater

 

or

 

Drinking

 

Water

 

Applications.

 

Since

 

there

 

is

 

no

 

membrane

 

or

 

electrolyte,

 

the

 

maintenance

 

cycle

 

is

 

dramatically

 

longer,

 

and

 

cleaning

 

tasks

 

are

 

usually

 

not

 

needed.

 

The

 

sensor

 

has

 

an

 

integrated

 

cleaning

 

system

 

using

 

polymeric

 

beads

 

the

 

cleans

 

all

 

the

 

electrodes,

 

including

 

the

 

pH

 

sensor

 

cartridge.

 

 

In

 

Wastewater

 

applications,

 

some

 

cleaning

 

may

 

be

 

necessary

 

when

 

visible

 

clogging

 

of

 

ports

 

occurs

 

due

 

to

 

biofouling

 

in

 

immersion

 

installations.

 

The

 

electrodes

 

remain

 

clean

 

but

 

the

 

flow

 

into

 

the

 

sensor

 

is

 

restricted

 

due

 

to

 

these

 

clogged

 

or

 

plugged

 

ports.

 

7.2.

 

Sensor Cleaning 

When

 

sensors

 

are

 

installed

 

in

 

applications

 

with

 

high

 

iron

 

concentrations

 

or

 

high

 

biofouling,

 

a

 

buildup

 

can

 

occur

 

on

 

the

 

end

 

of

 

the

 

sensor

 

(strainer)

 

that

 

may

 

restrict

 

flow

 

through

 

the

 

sensor.

 

In

 

these

 

applications,

 

the

 

sensor

 

should

 

be

 

inspected

 

every

 

two

 

weeks.

 

If

 

build

 

up

 

is

 

present,

 

follow

 

one

 

of

 

these

 

two

 

following

 

procedures:

  

 

Option

 

1:

 

Metabisulfite

 

then

 

the

 

sensor

 

should

 

be

 

immersed

 

for

 

one

 

minute

 

in

 

a

 

solution

 

of

 

Potassium

 

Metabisulfite

 

(LD

 

Carlson

 

6012A).

 

Follow

 

manufacturers

 

recommendation

 

for

 

solution

 

concentration.

 

The

 

sensor

 

should

 

be

 

operated

 

during

 

the

 

cleaning

 

step,

 

rinsed,

 

powered

 

off,

 

then

 

returned

 

to

 

service.

 

 

Option

 

2:

 

Muriatic

 

Acid

 

To

 

acid

 

clean

 

the

 

sensor,

 

remove

 

the

 

sensor

 

from

 

process,

 

rinse

 

to

 

remove

 

any

 

residual

 

chlorine.

 

1.

 

Select

 

a

 

small

 

container

 

and

 

fill

 

with

 

2”

 

of

 

tap

 

water

 

and

 

immerse

 

the

 

sensor.

 

2.

 

Power

 

the

 

sensor

 

on

 

and

 

tap

 

lightly

 

to

 

remove

 

air

 

and

 

rinse

 

the

 

sensor.

 

3.

 

Drain

 

and

 

refill

 

the

 

small

 

container

 

with

 

1:9

 

parts

 

muriatic

 

acid

 

to

 

water.

 

4.

 

Place

 

the

 

sensor

 

in

 

the

 

container,

 

power

 

on

 

and

 

clear

 

air

 

as

 

in

 

step

 

2.

 

5.

 

Run

 

the

 

sensor

 

for

 

30

 

to

 

60

 

seconds,

 

remove

 

the

 

sensor,

 

and

 

rinse

 

as

 

in

 

Step

 

2.

 

Do

 

not

 

leave

 

the

 

sensor

 

in

 

this

 

solution

 

for

 

more

 

than

 

60

 

seconds.

 

6.

 

Return

 

the

 

sensor

 

to

 

the

 

process

 

and

 

check

 

calibration.

 

 

WARNING:

 

ACIDS

 

ARE

 

HAZARDOUS.

 

Always

 

wear

 

eye

 

and

 

skin

 

protection

 

when

 

handling.

 

Follow

 

all

 

Material

 

Safety

 

Data

 

Sheet

 

recommendations.

 

 

7.3.

 

Flow Cell Maintenance 

The

 

maintenance

 

on

 

the

 

flow

 

cell

 

is

 

simple

 

cleaning.

 

The

 

flow

 

cell

 

is

 

clear

 

to

 

make

 

examination

 

of

 

the

 

condition

 

of

 

the

 

sensor

 

easier

 

without

 

interfering

 

with

 

operations.

 

The

 

flow

 

cell

 

may

 

be

 

cleaned

 

by

 

wiping

 

or

 

by

 

washing

 

with

 

detergents

 

or

 

diluted

 

acids.

 

Do

 

not

 

try

 

to

 

clean

 

with

 

solvents

 

as

 

the

 

acrylic

 

may

 

craze

 

or

 

crack.

 

Change

 

the

 

O

ring

 

annually

 

if

 

worn

 

or

 

still.

 

The

 

O

ring

 

can

 

be

 

lubricated

 

with

 

silicone

 

grease.

 

Replace

 

the

 

tubing

 

at

 

least

 

once

 

a

 

year.

 

Replace

 

brittle,

 

cracked

 

or

 

leaky

 

tubing

 

immediately.

 

Replace

 

old,

 

discolored,

 

fouled

 

or

 

damaged

 

tubing

 

with

 

tubing

 

of

 

the

 

same

 

diameter.

 

 
 
 

7.4.

 

Annual Maintenance 

7.4.1

 

Wear Part Replacement 

This

 

sec on

 

is

 

dedicated

 

to

 

the

 

vital

 

prac ce

 

of

 

replacing

 

wear

 

parts

 

in

 

Chlorine

 

Analyzers.

 

Wear

 

parts,

 

such

 

as

 

gaskets,

 

bearings,

 

impellers,

 

and

 

wear

 

rings,

 

undergo

 

natural

 

wear

 

and

 

tear

 

during

 

the

 

opera on

 

of

 

these

 

analyzers.

 

By

 

understanding

 

when

 

and

 

how

 

to

 

replace

 

these

 

components,

 

users

 

can

 

uphold

 

the

 

accuracy

 

of

 

their

 

measurements,

 

prolong

 

the

 

sensor's

 

service

 

life,

 

and

 

ensure

 

the

 

e

ciency

 

of

 

water

 

treatment

 

processes.

 

Summary of Contents for MP5A

Page 1: ...ble Diamond Pkwy Suite B10 Reno NV 89521 Phone 1 775 832 0495 www halogensys com info halogensys com GETTING TECHNICAL ASSISTANCE To get assistance for product installation or commissioning contact Ha...

Page 2: ...Options 15 6 1 Display Options 15 6 2 Output Configuration 15 6 3 Chlorine set up 4 20 mA 15 7 Maintenance 20 7 1 Wastewater Maintenance 20 7 2 Sensor Cleaning 20 7 3 Flow Cell Maintenance 20 7 4 Annu...

Page 3: ...nd commissioning If air is detected the sensor pump will cycle on and off three times to remove air from the sensor and may delay the first cycle completion by up to 20 seconds Figure 1 Start up scree...

Page 4: ...00 S Pressure limit 10 bar 145 psi Temperature 1 to 55 C Sample Compensation Automatic Factory calibration Yes Power consumption 24VDC 10 at 50mA 200 mA startup maximum Communication Modbus RTU RS485...

Page 5: ...ther than that specified in this manual Use of hazard information DANGER Indicates a potentially or imminently hazardous situation which if not avoided will result in death or serious injury WARNING I...

Page 6: ...20 mA Outputs 3 To a PC with an USB RS485 adapter The sensor is installed in a clear acrylic flow cell designed to purge air The inlet pressure should be limited to 60 PSI The flow rate can be from 0...

Page 7: ...he sensor facing up 3 Seat the sensor until the flange is flush with the top of the flow cell 4 Thread the collar down until it stops Hand tighten only 4 2 1Flow Cell Orientation Note that the orienta...

Page 8: ...1 1 Chlorine Negative Brown 2 1 PSU Positive or Conductivity Green 3 2 PSU Negative or Conductivity Yellow 4 2 Temperature Positive Gray 5 3 Temperature Negative Pink 6 3 pH Positive Blue 7 4 pH Nega...

Page 9: ...mm Slow Blow 3 5 Amp Labeled on the printed wiring board F1 and F2 5 Operation The start up screen will appear after the first cycle Figure 7 The display will take one minute to update The system cali...

Page 10: ...llows 1 Calibration check upon initial installation and normal process conditions 2 Check calibration at least every six months 3 If the value deviates by more than 20 then recalibrate the sensor Sinc...

Page 11: ...indicator will appear in the upper right area of the screen This indicates that the sensor is updating and will display the new value after it completes the cycle THE OLD VALUE WILL BE DISPLAYED UNTIL...

Page 12: ...ensor settings and calibrations have been changed and the sensor is not measuring correctly or has persistent error codes that cannot be resolved then restoring the default calibrations may resolve th...

Page 13: ...e following screen will appear You will need to get the Key Code from your supervisor or have a qualified person perform the calibration operation The Key Code consist of a series of keys pressed in o...

Page 14: ...ion is very stable If calibration is required the following procedure will adjust the pH reading 5 4 1Known Value To adjust the Known pH value follow these steps Press MENU button then select CALIBRAT...

Page 15: ...t 4 pH 5 to 12 Usually an integrator will want to match output settings to the units and scaling used on the PLC The most common changes will be detailed below The following example will change the ch...

Page 16: ...ple will change the output range from pH 5 to pH 12 to pH 0 to pH14 Use and use the keypress string below to get to the menu needed MENU 4 20 MA Options Set 4 20MA Range 4 20 PH DOWN DOWN DOWN SELECT...

Page 17: ...to set up the temperature 4 20 mA Output Change 4 20 mA Scaling Temp Settings Min Offset Max ppm Range span Disp Temp Default F 32 122 90 YES Desired C 0 40 40 NO Example 1 Set 4 20 mA Span from 32 F...

Page 18: ...less the OFFSET a Set Range to 122 32 90 090 Menu sequence will be MENU 4 20 MA Options Set 4 20MA Range DOWN DOWN SELECT Press DOWN arrow to decrement each digit to 90 then press SELECT to accept Fig...

Page 19: ...t 2 from PSU to Cond the label will not change to Cond even the 4 20 will be set for this output Be sure to adjust the high level to max conductivity level Scroll down to reach the ORP and CND lines A...

Page 20: ...remove the sensor from process rinse to remove any residual chlorine 1 Select a small container and fill with 2 of tap water and immerse the sensor 2 Power the sensor on and tap lightly to remove air...

Page 21: ...install new wear parts Remove parts from the replacement wear kit Install new wear ring into wear ring groove Install new impeller into impeller well Place 15 cleaning beads into the sensor end cover...

Page 22: ...ess in pH sensor Figure 16 and Figure 17 PN TOOL 0001 2 Rotate sensor counterclockwise CCW to unthread it Figure 18 3 Remove the old pH sensor and replace it with the new one Figure 19 4 Upon Startup...

Page 23: ...and correct 6 If the sensor still does not communicate then replace it with a new sensor Figure 20 Screen indicates no communication with sensor no sensor address found 8 2 Level One Problem Codes an...

Page 24: ...y be inaccurate Check process Replace sensor if TRO level is not accurate 37 pH low 5 Verify pH with handheld meter If sensor is not accurate replace the pH sensor 38 pH high 10 Verify pH with handhel...

Page 25: ...e take a photo of this screen To display this screen press the DOWN Arrow button from the main display This will bring up the screen below Press the Down Arrow again to display the next set of values...

Page 26: ...he sampling stream or the sensor orientation may be wrong Sensor is covered by brown slime or other contaminants Clean sensor using one of the two methods described in the Maintenance Section in the A...

Page 27: ...appropriate Accessory that you need for your installation Select Cable length if you need more than the Standard 1 5 M 5 length Select the Display type that you need There are options for 24VDC or 120...

Page 28: ...r Connection D01 Cable Gland PG21 24VDC Low Voltage Display for SCADA or Solar Power with 2 outputs 2 4 20 mA Cable Gland PG21 Cable Gland PG21 D01 4 Low Voltage Display for SCADA or Solar Power with...

Page 29: ...orine in the Display for up to 10 minutes This allows time after pulling the water sample for the DPD reaction 10 minutes The stored signal value will be used for the calibration of the sensor This is...

Page 30: ...o calibrate 1 2 3 or 4 using the up and down arrows 2 Connect an ammeter to output leads to read the live current of the output to be calibrated 3 Set display output to 4 mA by hitting select on Set o...

Page 31: ...eplacing a pH sensor using pH 7 and pH 10 buffers You can obtain these from any laboratory supply 1 Place sensor in 7 Buffer solution Navigate to Save Buffer 7 menu item Menu Calibration pH Buffer Cal...

Page 32: ...emperature Calibration Activate the Long Menu by holding Select and for 3 seconds Navigate to MENU Calibration Temp Enter desired temperature in Celsius Press Save Now 11 5 Changing Modbus Options MEN...

Page 33: ...it is sometimes useful to view the log file entries Log file entries are not timestamped but are recorded continuously in a loop There is a maximum of 65 000 entries The last log entries appear first...

Page 34: ...t shown ORP Oxidation Reduction Potential if enabled in Configuration Err Error Code See Troubleshooting section AUX Indicates if air is present Values about 2100 are usually due to air Can also indic...

Page 35: ...s several options to do this under Diagnostics The following steps may be used to adjust the slope and zero values to adjust the readings without going through the entire calibration process Figure 23...

Page 36: ...ual measurements of the sensor To disable 0 00 precision on the chlorine reading 0 01 scroll down to the Dsp 01 and put a dash next to this measurement If Conductivity is enabled PSU will not display...

Page 37: ...HL Menu Delayed Chlorine Cal Take sample DPD Level Save value Success Limit reached hi or low e14 CHL Zero Is sensor in Zero Y N Save Value Success Failed make sure in zero nnn nA e13 Fast Chlorine Ca...

Page 38: ...You Sure Y Restarting Configuration Measurement Options Do pH Do ORP Do PS EOC Hld Display Options Disp pH Disp ORP Dsp CND Dsp PSU Dsp 01 Key Cde Deg F Backlite 4 20 mA Options Set 4 20MA Span Upper...

Page 39: ...lope Edit Offset Logs View last log Erase logs Clear Errors 13 Other Installation Methods 13 1 Pipe Installation Locate the sensor in a straight length of pipe at least 1 5 times the pipe diameter bot...

Page 40: ...nd Maintenance Manual ________________________________________________________________________________________________________________ Copyright 2023 2024 Halogen Systems Inc All rights reserved world...

Page 41: ...dry area preferably at eye level using the tabs on the enclosure The side with the single knockout should face downward Figure 7 D 01 Display enclosure should be installed single knockout facing downw...

Page 42: ...ll the sensor using a 45 1 NPT thread FIP by socket adapter on the end of the sensor 3 Position the sensor so that the outlet port it oriented vertically to purge air with the 45 fitting 4 Run the sen...

Page 43: ...tube or pipe for mounting the components 13 4 2 Assembly 1 Follow instructions to assemble panel and enclosure on a 1 Pipe or tube not included 2 Pick a convenient spot for the Display 3 Wire a two w...

Page 44: ...y misuse neglect accident or improper application or installation Damage caused by any repair or attempted repair not authorized by the Halogen Systems Any product not used in accordance with the inst...

Page 45: ...tic and Semi Automatic Laboratory Equipment for Analysis and other Purposes UL 61010 2 081 2019 Ed 3 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Part 1 Gene...

Reviews: