MP5A Operation and Maintenance Manual
________________________________________________________________________________________________________________
Copyright
©
2023
‐
2024
Halogen
Systems,
Inc.
All
rights
reserved
worldwide.
Rev.:
4.02
14
Entering
the
Key
Code
will
unlock
the
calibration
menu
and
Sections
6
and
7
can
be
followed
to
adjust
calibrations.
5.3.
Key Code
5.4.
pH Calibration
The
MP5
pH
sensor
uses
a
no
‐
maintenance
solid
gel
in
its
pH
sensor
cartridge.
It
also
uses
DryGlass
which
is
stored
dry
and
wets
out
within
2
minutes
after
water
immersion.
The
Self
‐
Cleaning
function
keeps
the
glass
and
the
porous
junction
clean.
As
a
result,
the
calibration
is
very
stable.
If
calibration
is
required,
the
following
procedure
will
adjust
the
pH
reading.
5.4.1
Known Value
To
adjust
the
Known
pH
value,
follow
these
steps:
Press
MENU
button,
then
select
CALIBRATE
using
DOWN
arrow,
then
SELECT
Select
Known
pH
and
cursor
will
move
to
pH,
adjust
the
sensor
value
to
the
measured
reading.
It
will
take
one
cycle
to
reflect
the
new
calibration.
5.5.
Entering a Manual Fixed pH Value
If
your
pH
does
not
change,
you
can
select
a
fixed
pH
that
the
sensor
will
use
for
pH
compentation.
Enable
this
in
Measurement
Options
11.10
The
manual
level
to
be
used
is
entered
in
the
pH
calibration
screen.
C A L I B R A T E
* * * *
E N T E R
K E Y
C O D E
Enter
Key
Code
C H L
> p H
C a l i b r a t i o n
d e f a u l t
Select
pH
> S i n g l e
P o i n t
C a l
E n t e r
M a n u a l
p H
Calibrate
pH
> K n o w n
p H
7 .
6 4
P H Z
0 .
0
‐
5 9 .
2
Calibrate
pH