5: Operating the JAZ-INDY Module
42
013-RD000-000-02-201510
Four single-ended voltage inputs and the two differential inputs
4-20mA transmitter
4-20mA receiver
Calibrating the Voltage Outputs
The JAZ-INDY module provides four analog voltage outputs on its 26-pin connector (outputs 14 are
mapped to pins 5, 14, 23, and 4 respectively, where pins 2, 12, or 15 can be used as analog ground).
►
Procedure
1.
Connect an external, calibrated voltmeter (providing at least 5 total digits resolution) to each
output pin to calibrate it. No load should be connected to the output pin; it should be connected
to a voltmeter and nothing else.
2.
Set the output voltage (in counts) until the voltmeter reads as close as possible to -4.5V. Record
this value to be used later.
3.
Set the output voltage (in counts) until the voltmeter reads as close as possible to 0.0V. Record
this value to be used later.
4.
Set the output voltage (in counts) until the voltmeter reads as close as possible to 4.5V. Record
this value to be used later.
5.
Using the Jaz menu or OceanView software, select the voltage output to be calibrated:
a.
Select Minimum, then change the voltage to -4.5 and set the counts to the value you
measured in Step 2.
b.
Select Zero and enter the counts measured in Step 3.
c.
Select Maximum, then change the voltage to 4.5 and set the counts to the value you measured
in Step 4.
6.
Repeat this procedure for each of the output channels.
Calibrating the Voltage Inputs
Once the voltage outputs are calibrated, they can be used to calibrate the voltage inputs. These can be
found on pins 3, 13, 22, and 21. Note that the output voltage may sag slightly when connected to the
voltage inputs, so it is necessary to keep the voltmeter connected to the joined input and output pins to
compensate.
►
Procedure
1.
Set the output voltage (in counts) until the voltmeter reads as close as possible to -4.5V.
2. Using the Jaz menu or OceanView software (see
Calibrating the 4-20mA Transmitter
), select the
voltage input to be calibrated:
a.
Select Minimum, and change the voltage to -4.5V. Then select Sample at -4.50V. This will
update the counts automatically.
Summary of Contents for Ocean Optics JAZ
Page 6: ...Table of Contents iv 013 RD000 000 02 201510 ...
Page 10: ...About This Manual viii 013 RD000 000 02 201510 ...
Page 18: ...1 Introduction 8 013 RD000 000 02 201510 ...
Page 26: ...3 Operating Jaz 16 013 RD000 000 02 201510 Jaz Main Menu for DPUs dated January 1 2008 ...
Page 28: ...3 Operating Jaz 18 013 RD000 000 02 201510 Jaz Main Menu for DPUs dated after January 1 2008 ...
Page 31: ...4 Operating Jaz 013 RD000 000 02 201510 21 for DPUs after January 1 2008 ...
Page 35: ...4 Operating the JAZ PX Module 013 RD000 000 02 201510 25 JAZ PX Menu JAZ PX Menu Tree ...
Page 36: ...4 Operating the JAZ PX Module 26 013 RD000 000 02 201510 ...
Page 44: ...4 Operating the JAZ PX Module 34 013 RD000 000 02 201510 ...
Page 46: ...5 Operating the JAZ INDY Module 36 013 RD000 000 02 201510 JAZ INDY Menu ...
Page 47: ...5 Operating the JAZ INDY Module 013 RD000 000 02 201510 37 ...
Page 48: ...5 Operating the JAZ INDY Module 38 013 RD000 000 02 201510 ...
Page 49: ...5 Operating the JAZ INDY Module 013 RD000 000 02 201510 39 ...
Page 54: ...5 Operating the JAZ INDY Module 44 013 RD000 000 02 201510 ...
Page 60: ...6 Maintenance 50 013 RD000 000 02 201510 ...