4: Operating the JAZ-PX Module
013-RD000-000-02-201510
33
Synchronizing a Spectrometer Module to a JAZ-PX Module (recommended)
►
Procedure
1.
Getting Started using JAZ-PX with OceanView
to add Trigger Select and Strobe/Lamp
Enable to the Main Controls if they are not already there.
2.
Change the
Trigger Select Source
to
Internal
and set your
Trigger Period
to
10,000
µ
s as a
starting point. The lamp will now be controlled by OceanView software (via the Strobe/Lamp
Enable control).
3.
Scroll down to
Strobe/Lamp Enable
and check the box to turn the lamp on.
4.
Adjust the integration time in increments of 10 ms (Internal Trigger Period) to ensure that the
signal does not vary with each acquisition. It is also helpful to set the Scans to Average to
1
so
that the signal appears stable. After stability is set, then increase Scans to Average as desired.
Note
It is important to synchronize the lamp's flash rate (internal trigger period) with the
integration time. For instance, if the trigger period is 10 ms and the integration time is 60
ms, there would be 6 full flashes for each acquisition (consistent light each
measurement). Results will be inconsistent if the amount of light per acquisition varies (if
the times are out of sync).
Summary of Contents for Ocean Optics JAZ
Page 6: ...Table of Contents iv 013 RD000 000 02 201510 ...
Page 10: ...About This Manual viii 013 RD000 000 02 201510 ...
Page 18: ...1 Introduction 8 013 RD000 000 02 201510 ...
Page 26: ...3 Operating Jaz 16 013 RD000 000 02 201510 Jaz Main Menu for DPUs dated January 1 2008 ...
Page 28: ...3 Operating Jaz 18 013 RD000 000 02 201510 Jaz Main Menu for DPUs dated after January 1 2008 ...
Page 31: ...4 Operating Jaz 013 RD000 000 02 201510 21 for DPUs after January 1 2008 ...
Page 35: ...4 Operating the JAZ PX Module 013 RD000 000 02 201510 25 JAZ PX Menu JAZ PX Menu Tree ...
Page 36: ...4 Operating the JAZ PX Module 26 013 RD000 000 02 201510 ...
Page 44: ...4 Operating the JAZ PX Module 34 013 RD000 000 02 201510 ...
Page 46: ...5 Operating the JAZ INDY Module 36 013 RD000 000 02 201510 JAZ INDY Menu ...
Page 47: ...5 Operating the JAZ INDY Module 013 RD000 000 02 201510 37 ...
Page 48: ...5 Operating the JAZ INDY Module 38 013 RD000 000 02 201510 ...
Page 49: ...5 Operating the JAZ INDY Module 013 RD000 000 02 201510 39 ...
Page 54: ...5 Operating the JAZ INDY Module 44 013 RD000 000 02 201510 ...
Page 60: ...6 Maintenance 50 013 RD000 000 02 201510 ...