Chapter 7, Troubleshooting and Maintenance
85
Haemonetics
®
TEG
®
6s User Manual
P/N 115191-IE, Manual revision: AB
1016
Bubble detected in
channel.
Bubbles have been
detected in one or more
channels.
Remove the cartridge and
rerun the test using a new
cartridge. Verify that the
quantity of blood sample is
sufficient for the test. If the
problem persists, contact
the local Haemonetics
representative.
1030
Pressure leak in
instrument
There was a pressure leak
in the analyzer.
Contact the local
Haemonetics
representative.
1031
Pressure leak in cartridge
There was a pressure leak
in the cartridge.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
1032
Vacuum leak in instrument
There was a vacuum leak
in the analyzer.
Contact the local
Haemonetics
representative.
1033
Vacuum leak in cartridge
There was a vacuum leak
in the cartridge.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
1034
Regulated vacuum error
in cartridge
The regulated vacuum
was out of range when
the cartridge was
clamped.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
1037
Regulated pressure error
in instrument
The regulated vacuum
was out of range when
the cartridge was
not
clamped.
Contact the local
Haemonetics
representative.
2001
Front panel LED failure
The front panel LED failed. The analyzer will continue
to operate correctly.
Report the failure to the
local Haemonetics
representative.
Error Code
LCD Display Message
Probable Cause
Recommended Action