84
Chapter 7, Troubleshooting and Maintenance
P/N 115191-IE, Manual revision: AB
Haemonetics
®
TEG
®
6s User Manual
Figure 7-3, Critical error screen
Error message
table
The following table provides some examples of errors and alerts that you may
encounter and includes the error code, display message, probable cause, and
recommended solution. For error codes that are displayed on the analyzer but
are not included in the table, please contact the local Haemonetics
representative for assistance.
Error Code
LCD Display Message
Probable Cause
Recommended Action
1000
The instrument has
unexpectedly been reset.
The analyzer was
unexpectedly reset while
idle.
Turn the power off and
then back on to reboot the
analyzer. If the problem
persists, contact the local
Haemonetics
representative.
1005
Power-On-Self-Test
optics error.
During POST, one or
more photodetectors are
reading abnormally high
or low, or it was not
possible to calibrate the
excitation LED(s) as
desired.
Turn the power off and
then back on to reboot the
analyzer. If the problem
persists, contact the local
Haemonetics
representative.
1010
Power-On-Self-Test
vacuum leak error.
A vacuum leak was
detected during POST.
Turn the power off and
then back on to reboot the
analyzer. If the problem
persists, contact the local
Haemonetics
representative.
Summary of Contents for TEG 6s
Page 1: ...User manual HAEMONETICS S A Signy Centre Rue des Fl ch res 6 1274 Signy Centre Switzerland...
Page 4: ......
Page 8: ......
Page 16: ......
Page 28: ......
Page 51: ...Chapter 4 Getting Started 51 Haemonetics TEG 6s User Manual P N 115191 IE Manual revision AB...
Page 67: ......
Page 79: ......