Chapter 6, Quality Control
71
Haemonetics
®
TEG
®
6s User Manual
P/N 115191-IE, Manual revision: AB
5. Prepare the QC sample according to the manufacturer’s instructions.
6. Pipette the QC sample into the cartridge sample port, filling up to or
above the line marked on the cartridge, and then touch
next
.
7. View the test results on the results and/or tracing screens.
8. When prompted to remove the cartridge, dispose of it properly.
9. Touch
done
to return to the main menu.
Detailed guide
for running a
QC test
1. From the
Home
screen, touch
new qc
.
Figure 6-1, Home screen – touch
new qc
2. Remove the desired QC cartridge from its sealed pouch.
3. Insert the cartridge into the slot, as indicated on the
Preparing Test
–
Insert cartridge
screen, with the bar code on the left side.
Note: Only a Haemonetics QC cartridge can be inserted into the cartridge
slot. Check the label to be sure you are using the intended QC assay.