Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3Vrms
150kHz to 80MHz,
1kHZ 80%AM
Modulation
6Vrms in ISM
bands(I/O cables< 3m
excluded)
Patient coupled ports
tested with current
clamp
3V/m,
80Mhz to 2,7GHz,
1kHz 80%AM
modulation
Table-9
(IEC60601-1-2:2014)
3Vrms
150kHz to
80MHz,
1kHZ 80%AM
Modulation
6Vrms in ISM
bands(I/O
cables< 3m
excluded)
Patient
coupled ports
tested with
current clamp
3V/m,
80Mhz to
2,7GHz, 1kHz
80%AM
modulation
Table-9
(IEC60601-1-
2:2014)
d
= 1,2√
P
80 to 800MHz
d
= 2,3√
P
800MHz to 2,5GHz
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d
is the recommended separation
distance in meters (m).
Field strength from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the
compliance level in each frequency
range.
b
Interference may occur in the vicinity of
the equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the ES-100V3 is used exceeds the applicable RF compliance level above, the
ES-100V3 should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the ES-100V3.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.