26
Bayonet
10MHz:
NRP-10H*
20MHz:
NRP-20H*
Flexible
10MHz:
NRP-10HF*
20MHz:
NRP-20H1NF*
Single use
10MHz:
NDP-10H*
20MHz:
NDP-20H*
*: Except European Union Countries.
5-3-2. Sterilization
Only sterilizable probes can be sterilized. Do not sterilize other type of
probes including amplifiers as well as main unit.
Warnings:
Sterilizable probes are not sterilized before shipment.
They must be sterilized before use as follows:
Sterilization limits:
All sterilizable probes except ACP and FDP probe: Up to 50 times
ACP probe: Up to 5 times by steam autoclave
FDP probe: Up to 5 times
Note: Do not exceed sterilization limits or it may cause damage to probes.
Cautions:
1.
Except ACP probe, do not sterilize probes by steam autoclave nor put
them in washer disinfector or it will damage probes.
2.
ACP probe should be sterilized by steam autoclave as described in
section "Instructions for sterilization" below.
Instructions for sterilization:
Point of preparation:
No particular requirements.
Preparation for cleaning: No particular requirements.
Cleaning:
Automated: Do not do automated cleaning of probes other than ACP
probe.