42
(4) Ask the patient to flex the foot specified
number on COUNT following the foot
animation and beep on Bidop. The
exercise should be forceful, especially
when lifting the foot upward.
(5) After flexing, instruct the patient to relax
the foot and avoid all movement.
(6) The test is completed when the
waveform returns to the baseline and
Bidop will automatically freeze the
waveform and calculate recovery times.
Note: "1/2" is the half recovery time for returning to 50% of refilling amplitude
where middle vertical dotted line is shown.
(7) If you wish to store the data on the
memory, see Ҥ 4-2-a. MEMORY -
STORE”.
(8) Press the probe button to get out of the
freeze mode.