28
Connect the probe to main unit and make sure if you
hear Doppler sounds properly when you rub probe tip.
Packaging:
No particular requirements.
Storage:
No particular requirements.
Manufacturer contact:
Hadeco, Inc.
2-7-11 Arima, Miyamae-ku, Kawasaki, 216-0003, Japan
Tel : +81-44-877-4361
Fax : +81-44-855-7301
The instructions provided above have been validated by the medical device
manufacturer as being CAPABLE of preparing a medical device for reuse. It
remains the responsibility of the processor to ensure that the processing as
actually performed using equipment, materials and personal in the processing
facility achieve the desired result. This requires validation and routine
monitoring of process. Likewise any deviation by the processor from the
instruction provided should be evaluated for effectiveness and potential
adverse consequences.