21
f. TIME (Time scale)
(1) Press Shuttle Button to change the time scale as follows:
• Normal:
For arteries (2.56 sec./screen)
• Slow:
For veins (25.6 sec./screen)
g. DISP (DISP, OTHERS
– DISP, DISP/BEEP)
(1) Press Shuttle Button to change the Display mode as follows:
<DISP>
• WAVE:
Waveform
• DATA:
Numerical data
• BEEP:
2MHz BEEP mode
<DISP/BEEP>
• WAVE
Waveform for 2MHz BEEP mode
• DATA
Numerical data for 2MHz BEEP mode
Note: On heart rate mode, display mode cannot be changed when on Freeze
mode. When on heart rate measurement mode with WAVE mode, turning
Shuttle button up & down to restart the monitoring.
h. SOUND (Beep sound for HR)
Set SOUND mode for ON to activate beep sounds when heart rate gets out
of upper and lower limits during the test except 1
st
30 sec.
(1) Press Shuttle Button to change beep sound ON / OFF.
i. UPPER (Upper limit for HR)
In case heart rate exceeds the upper limit during the test except
1
st
30 sec.,
LCD will start flashing.
(1) Turn Shuttle Button up and down to select the upper limit in 5 BPM
steps and press the button to set it.
j. LOWER (Lower limit for HR)
In case heart rate gets below lower limit during the test except 1
st
30sec.,
LCD will start flashing.