14.
Two options are available during the cell conditioning procedure:
Option Description
Stop
Aborts the measurement and no results are calculated. If selected during the
Replicate Sample
option, then all data in the series is lost.
Skip
Stops the cell conditioning and goes directly to the next step in the procedure. Results are then
calculated from measurement data available before
Skip
is selected. Results can be less accurate
using this option.
15.
Weigh the sample in its container and write down the result.
16.
When the cell conditioning is complete add the sample. Remove the sample stopper and put the
correct amount of sample into the beaker with an applicable utensil. Make sure that all of the
sample is put into the center of the beaker and that none of it spills onto the walls of the beaker.
17.
Put the sample stopper back into the sensor holder. Make sure that the sample stopper fits tightly
into the sensor holder.
18.
Weigh the empty sample container and subtract this value from the weight of the sample in its
container (refer to step
). The result is the exact weight of the sample analyzed. Write down
this value.
19.
Push
Start
to start the moisture content analysis. Measurement data shows on the display.
Note: The moisture content analysis will start automatically after the addition of sample if the Injection
Autodetect parameter is set to Yes.
20.
If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
Note: Changing the default stirring speed can effect the measurement result.
21.
Two options are available during the measurement procedure:
Option Description
Stop
Aborts the measurement and no results are calculated. If selected during the
Replicate Sample
option, then all data in the series is lost.
Skip
Stops the measurement and goes directly to the next step in the procedure. Results are then
calculated from measurement data available before
Skip
is selected. Results can be less accurate
using this option.
22.
When the measurement is complete, enter the exact weight of the sample analyzed (refer to step
). If necessary, push the arrow keys to change the value shown on screen. Push
Select
to
accept the value.
23.
If the
Cell standby
option has been selected (refer to
on page 18), the standby
operation starts again automatically.
24.
Push the arrow keys to see the different measurement views or push
Next
for the options that
follow:
Option
Description
Replicate sample
Use this option to start the same titration on the same sample. This is used to study the
repeatability by successively analyzing several parts of the same sample. At the end of
each measurement, a window shows the average value, the standard deviation and the
relative standard deviation.
New sample
Use this option to start the same titration on a new sample. No standard deviation or
relative standard deviation measurements will be done.
Cell management
Use this option to fill or empty the measurement cell using the pumps. Refer to
on page 25.
25.
Push
Exit
to go back to the home screen.
Syringe verification
Use the syringe verification option to examine the syringe accuracy based on the ISO 8655 standard.
English
23
Summary of Contents for TitraLab KF1121
Page 9: ...English 9...
Page 14: ...14 English...
Page 34: ...34 Deutsch...
Page 39: ...Deutsch 39...
Page 60: ...60 Italiano...
Page 65: ...Italiano 65...
Page 86: ...86 Fran ais...
Page 91: ...Fran ais 91...
Page 112: ...112 Espa ol...
Page 117: ...Espa ol 117...
Page 138: ...138 Portugu s...
Page 143: ...Portugu s 143...
Page 169: ...e tina 169...
Page 189: ...Dansk 189...
Page 194: ...194 Dansk...
Page 214: ...214 Nederlands...
Page 219: ...Nederlands 219...
Page 240: ...240 Polski...
Page 245: ...Polski 245...
Page 271: ...Svenska 271...
Page 290: ...290 Suomi...
Page 295: ...Suomi 295...
Page 309: ...ESD IECS 003 309...
Page 311: ...1 6 11 2 2 7 12 1 3 USB 1 8 13 4 9 5 10 1 KF1121 1 2 1 USB USB 311...
Page 312: ...USB USB USB USB 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 1 1 3 2 4 312...
Page 313: ...2 1 KF 6 2 KF PTFE 7 3 8 4 USB 9 3 x GL45 5 10 3x 313...
Page 314: ...15 6 in GFCI GFI 1 2 312 3 180 6 314...
Page 315: ...315...
Page 316: ...316...
Page 317: ...317...
Page 318: ...3 318...
Page 319: ...3 1 2 3 MSDS SDS 333 1 2 319...
Page 320: ...320...
Page 321: ...4 1 3 2 321...
Page 322: ...1 3 5 2 4 2 Settings Options 322...
Page 323: ...1 5 2 6 2 3 KF 7 4 8 2 2 USB 1 2 331 323...
Page 324: ...MSDS SDS 1 Settings 2 Select 324 3 Back 324...
Page 326: ...MSDS SDS a 1 2 3 USB 4 332 5 Calibration 6 Select 7 8 9 Start 10 11 326...
Page 333: ...300 C 572 F 4 333...
Page 341: ...Magyar 341...
Page 346: ...346 Magyar...
Page 367: ...Rom n 367...
Page 372: ...372 Rom n...
Page 392: ...392 T rk e...
Page 397: ...T rk e 397...
Page 417: ...Slovensk jazyk 417...
Page 422: ...422 Slovensk jazyk...
Page 441: ...Slovenski 441...
Page 446: ...446 Slovenski...
Page 465: ...Hrvatski 465...
Page 470: ...470 Hrvatski...
Page 484: ...IECS 003 A IECS 484...
Page 486: ...1 6 11 2 2 7 12 1 3 USB1 8 13 4 9 5 10 1 KF1121 1 2 USB USB USB 1 USB USB 486...
Page 487: ...USB 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 1 1 3 2 4 487...
Page 488: ...2 1 KF 6 2 KF PTFE 7 3 8 4 USB 9 3 x GL45 5 10 3x 488...
Page 489: ...15 cm AC GFCI GFI 1 2 486 3 180 6 489...
Page 490: ...490...
Page 491: ...491...
Page 492: ...492...
Page 493: ...3 493...
Page 494: ...3 1 2 3 MSDS SDS o 508 1 2 494...
Page 495: ...495...
Page 496: ...4 1 3 2 496...
Page 497: ...1 ON OFF 3 5 2 4 ON OFF 2 Settings Options 497...
Page 498: ...1 5 2 6 2 3 KF 7 4 8 2 2 USB 1 2 506 498...
Page 508: ...Empty cell pump 2 2 2 Stop pumps Fill cell Empty cell Fill cell Empty cell 3 Exit 300 C 4 508...
Page 509: ......