14.
When the cell conditioning is complete add the standard. Remove the sample stopper and put the
correct amount of standard into the beaker with an applicable utensil. Make sure that all of the
standard is put into the center of the beaker and that none of it spills onto the walls of the beaker.
15.
Put the sample stopper back into the sensor holder. Make sure that the sample stopper fits tightly
into the sensor holder.
16.
Weigh the empty sample container and subtract this value from the weight of the standard in its
container (refer to step
). The result is the exact weight of the sample analyzed. Write down
this value.
17.
Push
Start
to start the moisture content analysis. Calibration data shows on the display.
Note: The moisture content analysis will start automatically after the addition of sample if the Injection
Autodetect parameter is set to Yes.
18.
If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
Note: Changing the default stirring speed can effect the measurement result.
19.
Two options are available during the calibration procedure:
Option Description
Stop
Aborts the calibration and no results are calculated. If selected during the
Replicate Sample
option,
then all data in the series is lost.
Skip
Stops the calibration and goes directly to the next step in the procedure. Results are then calculated
from calibration data available before
Skip
is selected. Results can be less accurate using this
option.
20.
When the calibration is complete, enter the exact weight of the sample analyzed (refer to step
). If necessary, push the arrow keys to change the value shown on screen. Push
Select
to
accept the value.
21.
When the calibration is complete, push the left and right arrow keys to see the different calibration
views.
22.
Push
Reject
or
Continue
.
Option
Description
Reject
Rejects the calibration. Select
Cancel
to go back to the result display or
Confirm
to reject the
calibration. If this is the first calibration, select
Confirm
to reject the calibration and use the default
or previous calibration values. If this is a
Replicate Standard
calibration, select
Confirm
to reject
only the current calibration in the series.
Continue
Select from one of these options:
•
Replicate Standard
: Do the calibration again using the same standard
•
Save & Exit
: Keep the calibration results and exit the calibration procedure
•
Reject & Exit
: Reject the calibration results and use the default or previous calibration values,
and exit the calibration procedure
Get a sample measurement
Use this option to get sample measurements with one of the installed applications.
1.
Make sure that the glass beaker, the sensor holder and all related parts are clean and dry. Put a
magnetic stir bar into the beaker.
2.
Attach the beaker to the sensor holder. Make sure that the beaker fits tightly into the sensor
holder.
3.
Read the related "Application Note" from the USB applications key for more instructions.
4.
Fill the beaker with the applicable amount of solvent. Refer to the application note and
on page 25 to adapt the volume where applicable. Make sure that there is no
liquid spill. Make sure that the initial level is enough so that the sensor is correctly installed in the
sample. Do not put too much solvent in the beaker.
5.
From the home screen, select the measurement application, then push
Select
.
English
21
Summary of Contents for TitraLab KF1121
Page 9: ...English 9...
Page 14: ...14 English...
Page 34: ...34 Deutsch...
Page 39: ...Deutsch 39...
Page 60: ...60 Italiano...
Page 65: ...Italiano 65...
Page 86: ...86 Fran ais...
Page 91: ...Fran ais 91...
Page 112: ...112 Espa ol...
Page 117: ...Espa ol 117...
Page 138: ...138 Portugu s...
Page 143: ...Portugu s 143...
Page 169: ...e tina 169...
Page 189: ...Dansk 189...
Page 194: ...194 Dansk...
Page 214: ...214 Nederlands...
Page 219: ...Nederlands 219...
Page 240: ...240 Polski...
Page 245: ...Polski 245...
Page 271: ...Svenska 271...
Page 290: ...290 Suomi...
Page 295: ...Suomi 295...
Page 309: ...ESD IECS 003 309...
Page 311: ...1 6 11 2 2 7 12 1 3 USB 1 8 13 4 9 5 10 1 KF1121 1 2 1 USB USB 311...
Page 312: ...USB USB USB USB 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 1 1 3 2 4 312...
Page 313: ...2 1 KF 6 2 KF PTFE 7 3 8 4 USB 9 3 x GL45 5 10 3x 313...
Page 314: ...15 6 in GFCI GFI 1 2 312 3 180 6 314...
Page 315: ...315...
Page 316: ...316...
Page 317: ...317...
Page 318: ...3 318...
Page 319: ...3 1 2 3 MSDS SDS 333 1 2 319...
Page 320: ...320...
Page 321: ...4 1 3 2 321...
Page 322: ...1 3 5 2 4 2 Settings Options 322...
Page 323: ...1 5 2 6 2 3 KF 7 4 8 2 2 USB 1 2 331 323...
Page 324: ...MSDS SDS 1 Settings 2 Select 324 3 Back 324...
Page 326: ...MSDS SDS a 1 2 3 USB 4 332 5 Calibration 6 Select 7 8 9 Start 10 11 326...
Page 333: ...300 C 572 F 4 333...
Page 341: ...Magyar 341...
Page 346: ...346 Magyar...
Page 367: ...Rom n 367...
Page 372: ...372 Rom n...
Page 392: ...392 T rk e...
Page 397: ...T rk e 397...
Page 417: ...Slovensk jazyk 417...
Page 422: ...422 Slovensk jazyk...
Page 441: ...Slovenski 441...
Page 446: ...446 Slovenski...
Page 465: ...Hrvatski 465...
Page 470: ...470 Hrvatski...
Page 484: ...IECS 003 A IECS 484...
Page 486: ...1 6 11 2 2 7 12 1 3 USB1 8 13 4 9 5 10 1 KF1121 1 2 USB USB USB 1 USB USB 486...
Page 487: ...USB 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 1 1 3 2 4 487...
Page 488: ...2 1 KF 6 2 KF PTFE 7 3 8 4 USB 9 3 x GL45 5 10 3x 488...
Page 489: ...15 cm AC GFCI GFI 1 2 486 3 180 6 489...
Page 490: ...490...
Page 491: ...491...
Page 492: ...492...
Page 493: ...3 493...
Page 494: ...3 1 2 3 MSDS SDS o 508 1 2 494...
Page 495: ...495...
Page 496: ...4 1 3 2 496...
Page 497: ...1 ON OFF 3 5 2 4 ON OFF 2 Settings Options 497...
Page 498: ...1 5 2 6 2 3 KF 7 4 8 2 2 USB 1 2 506 498...
Page 508: ...Empty cell pump 2 2 2 Stop pumps Fill cell Empty cell Fill cell Empty cell 3 Exit 300 C 4 508...
Page 509: ......