Section 9
Troubleshooting
Refer to the following tables for problem messages and possible causes and solutions.
Table 8 Measurement
Error/Warning
Description
Possible solution
Measurement unstable
The measurement from the sensor is not
stable
Make sure that there are no bubbles around
the sensor element or junction
Measurement out of range
The sample concentration is outside the
sensor measurement limits
Adjust the sample preparation to be inside the
sensor measurement range
Use a different sensor with a more applicable
measurement range
Temperature out of range
The sample temperature is outside the
sensor measurement limits
Adjust the sample temperature to be inside
the sensor measurement range
Use a different sensor with a more applicable
measurement range
Make sure that the sensor is operating
correctly
Cell cond. skipped
The cell conditioning has been stopped
manually by the user
Refer to the application note for best practices
and do the titration again
Analysis skipped
The titration has been stopped manually by
the user
Refer to the application note for best practices
and do the titration again
Equiv. point not found
The titration fails to determine the
equivalence point
Refer to the application note for best practices
and do the titration again
Adjust the equivalence point detection value
for the application (refer to
Contact technical support
Out of range
The computed result is outside the accepted
limits
Refer to the application note for best practices
and do the titration again
Adjust the result minimum and maximum
values for the application (refer to
Equation calc. error
An equation variable is unknown (e.g. the
ordinate of the equivalence point is unknown
or the volume of the equivalence point is
unknown)
Make sure that all the equation variables are
available (refer to the application note for the
equation description and values)
Titrant calibration has expired
A titration has been done with an expired
titrant calibration date
Calibrate the titrant and if necessary adjust
the calibration frequency (refer to the
application note for the recommended
frequency)
The titrant used by an
application has been
modified and this application
became incompatible with
applications already installed.
Please go to application
management to solve the
issue
The application data has been changed and
the application cannot be used
Remove the application or change the titrant
to a titrant that is already installed
System error
A system error causes the titration to stop
(e.g. a sensor has been disconnected or the
volume of reagent is too low)
Make sure that all the system elements are
connected and working correctly (refer to
on page 41)
47
Summary of Contents for TitraLab KF1000 Series
Page 1: ...DOC022 52 93082 TitraLab KF1000 series workstations User Manual 03 2019 Edition 9...
Page 2: ......
Page 6: ...Specifications 4...
Page 12: ...General information 10...
Page 20: ...Installation 18...
Page 30: ...Startup 28...
Page 40: ...Standard operations 38...
Page 54: ...Troubleshooting 52...
Page 60: ...Replacement parts and accessories 58...
Page 64: ...Change application parameters 62...
Page 66: ...Frequently asked questions 64...
Page 67: ......