Option
Description
Export data log
Exports measurement data from the system to an external device. Preview
data selection before it is exported. Make sure that an external device is
connected to the instrument (e.g., a USB key, external hard drive, etc.).
Delete data log
Removes measurement data from the system. Previews data selection
before it is removed.
3.
Specify the data selection parameters. Push the left and right arrow keys to make a
selection. Push the up and down arrow keys to select an option.
Option
Description
Result type
Sets the type of result available.
Application
Sets the available applications.
Date
Sets the date range.
Operator
Sets the available operators.
4.
If
View data log
was the selected option, push
View
to see the selected data.
•
Push the up and down arrow keys to select a line of data and push
Detail
to see
more data
•
If only one
Application
is selected push the up and down arrow keys to select a
line of data and push
Detail
to see more data, or push the left and right arrow
keys to see related graphs
5.
If
Export data log
or
Delete data log
was the selected option, push
Preview
to see
the selected data, then push
Export
or
Delete
to start the procedure.
7.6 Purge
Use this procedure to remove air bubbles from the system. Refer to
on page 27 for instructions.
7.7 Cell management
Use this option to fill or empty the measurement cell with the pumps. These procedures
stop automatically after a pre-determined time. The time remaining is shown against
Security time
on the display.
1.
From the home screen, select
Cell management
.
2.
Push the left and right arrow keys to make a selection.
Option
Description
Fill cell (pump
1)
Use this option to fill the measurement cell with reagent from pump 1.
Empty cell
(pump 2)
C A U T I O N
Chemical exposure hazard. Before selecting this option, examine the
level of liquid in the waste bottle. Make sure that there is enough space in
the bottle for all of the liquid from the measurement cell.
Use this option to empty the measurement cell with pump 2.
Stop pumps
Use this option to stop the
Fill cell
and
Empty cell
procedures.
Stir
Use this option to start or stop the magnetic stirrer. This option can also be
used when the
Fill cell
and
Empty cell
procedures are operating.
3.
Push
Exit
to go back to the home screen.
Standard operations
37
Summary of Contents for TitraLab KF1000 Series
Page 1: ...DOC022 52 93082 TitraLab KF1000 series workstations User Manual 03 2019 Edition 9...
Page 2: ......
Page 6: ...Specifications 4...
Page 12: ...General information 10...
Page 20: ...Installation 18...
Page 30: ...Startup 28...
Page 40: ...Standard operations 38...
Page 54: ...Troubleshooting 52...
Page 60: ...Replacement parts and accessories 58...
Page 64: ...Change application parameters 62...
Page 66: ...Frequently asked questions 64...
Page 67: ......