1.
Put an empty beaker on the balance. Refer to the illustrated steps that follow.
2.
For the fully automated mode, connect the balance to the instrument.
3.
Install the support stand near the balance. Put the pipette tip adapter above the
beaker on the balance.
4.
Connect the tube from the pipette holder to the syringe outlet port.
5.
Connect the tube from the water bottle to the syringe inlet port.
6.
On the instrument, manually enter the temperature. As an alternative, connect and
put a temperature sensor on the water bottle to get the temperature automatically.
Note: The temperature setting is necessary for the fully automated and partially automated
modes. The dispense only mode does not adjust the temperature.
Install accessories
20