Section 8
Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned
or repaired, contact the manufacturer.
8.1 Maintenance schedule
shows the recommended schedule of maintenance tasks. Facility requirements
and operating conditions may increase the frequency of some tasks.
Table 7 Maintenance schedule
Task
1 day 7 days 12 months
on page 39. Also do this procedure immediately after any liquid spills on
the instrument.
x
on page 40.
x
Regenerate the molecular sieve
on page 40
x
Examine all tube connections for leaks and damage. Replace as necessary. Refer to
Replacement parts and accessories
on page 53 for part numbers and to
on page 40.
x
x
Replace the Viton O-ring in the measurement cell. Refer to
Replacement parts and accessories
on page 53 for part numbers.
x
Replace syringe electro-valve block. Refer to
Replace the syringe electro-valve block
on page 40.
x
Examine the syringe connections and inlet and outlet tubes for leaks and damage. Replace as
necessary. Refer to
Replacement parts and accessories
on page 53 for part numbers and to
on page 40.
x
Replace the pump cassettes. Refer to
on page 42.
x
Examine the pump connections and inlet and outlet tubes for leaks and damage. Replace as
necessary. Refer to
Replacement parts and accessories
on page 53 for part numbers and to
on page 40.
x
Examine the bottle caps and connections for leaks and damage. Replace as necessary. Refer
to
Replacement parts and accessories
on page 53 for part numbers.
x
8.2 Clean the instrument
N O T I C E
Never use flammable or corrosive solvents to clean any part of the instrument. Use of these
solvents can degrade the environmental protection of the instrument and may void the warranty.
Clean the exterior surface with a moist cloth or with a mixture of water and mild detergent.
Dry with a soft cloth.
39
Summary of Contents for TitraLab KF1000 Series
Page 1: ...DOC022 52 93082 TitraLab KF1000 series workstations User Manual 03 2019 Edition 9...
Page 2: ......
Page 6: ...Specifications 4...
Page 12: ...General information 10...
Page 20: ...Installation 18...
Page 30: ...Startup 28...
Page 40: ...Standard operations 38...
Page 54: ...Troubleshooting 52...
Page 60: ...Replacement parts and accessories 58...
Page 64: ...Change application parameters 62...
Page 66: ...Frequently asked questions 64...
Page 67: ......